Error on the stopping power ratio of ERKODENT's mouthpiece for head and neck carbon ion radiotherapy treatment

Author:

Lee Sung Hyun1,Kanai Takayuki12,Souda Hikaru1,Miyasaka Yuya1,Chai Hongbo1,Ono Takuya1,Yamazawa Yoshifumi13,Suzuki Koji3,Sato Azusa4,Katsumata Masashi4,Iwai Takeo1

Affiliation:

1. Department of Heavy Particle Medical Science, Graduate School of Medical Science Yamagata University Iidanishi Yamagata Japan

2. Department of Radiation Oncology Tokyo Women's Medical University Shinjuku Tokyo Japan

3. Department of Radiology Yamagata University Hospital Iidanishi Yamagata Japan

4. Accelerator Engineering Corporation Inage Chiba Japan

Abstract

AbstractThe errors on the stopping power ratio (SPR) of mouthpiece samples from ERKODENT were evaluated. Erkoflex and Erkoloc‐pro from ERKODENT and samples that combined Erkoflex and Erkoloc‐pro were computed tomography (CT)‐scanned using head and neck (HN) protocol at the East Japan Heavy Ion Center (EJHIC), and the values were averaged to obtain the CT number. The integral depth dose of the Bragg curve with and without these samples was measured for 292.1, 180.9, and 118.8 MeV/u of the carbon‐ion pencil beam using an ionization chamber with concentric electrodes at the horizontal port of the EJHIC. The average value of the water equivalent length (WEL) of each sample was obtained from the difference between the range of the Bragg curve and the thickness of the sample. To calculate the difference between the theoretical and measured values, the theoretical CT number and SPR value of the sample were calculated using the stoichiometric calibration method. Compared with the Hounsfield unit (HU)‐SPR calibration curve used at the EJHIC, the SPR error on each measured and theoretical value was calculated. The WEL value of the mouthpiece sample had an error of approximately 3.5% in the HU‐SPR calibration curve. From this error, it was evaluated that for a mouthpiece with a thickness of 10 mm, a beam range error of approximately 0.4 mm can occur, and for a mouthpiece with a thickness of 30 mm, a beam range error of approximately 1 mm can occur. For a beam passing through the mouthpiece in HN treatment, it would be practical to consider a mouthpiece margin of 1 mm to avoid beam range errors if ions pass through the mouthpiece.

Publisher

Wiley

Subject

Radiology, Nuclear Medicine and imaging,Instrumentation,Radiation