Clinical assessment of a novel machine‐learning automated contouring tool for radiotherapy planning

Author:

Hu Yunfei1ORCID,Nguyen Huong2,Smith Claire2,Chen Tom3,Byrne Mikel4,Archibald‐Heeren Ben1,Rijken James5,Aland Trent2

Affiliation:

1. Icon Cancer Centre Concord Rusty Priest Building Concord Repatriation Hospital Concord NSW Australia

2. ICON Core Office South Brisbane QLD Australia

3. Icon Cancer Centre Springfield Cancer Care Centre Mater Private Hospital 30 Health Care Dr Springfield Queensland Australia

4. Icon Cancer Centre Wahroonga Sydney Adventist Hospital Sydney Australia

5. Icon Cancer Centre Windsor Gardens Windsor Gardens South Australia Australia

Abstract

AbstractContouring has become an increasingly important aspect of radiotherapy due to inverse planning. Several studies have suggested that the clinical implementation of automated contouring tools can reduce inter‐observer variation while increasing contouring efficiency, thereby improving the quality of radiotherapy treatment and reducing the time between simulation and treatment. In this study, a novel, commercial automated contouring tool based on machine learning, the AI‐Rad Companion Organs RT™ (AI‐Rad) software (Version VA31) (Siemens Healthineers, Munich, Germany), was assessed against both manually delineated contours and another commercially available automated contouring software, Varian Smart Segmentation™ (SS) (Version 16.0) (Varian, Palo Alto, CA, United States). The quality of contours generated by AI‐Rad in Head and Neck (H&N), Thorax, Breast, Male Pelvis (Pelvis_M), and Female Pelvis (Pevis_F) anatomical areas was evaluated both quantitatively and qualitatively using several metrics. A timing analysis was subsequently performed to explore potential time savings achieved by AI‐Rad. Results showed that most automated contours generated by AI‐Rad were not only clinically acceptable and required minimal editing, but also superior in quality to contours generated by SS in multiple structures. In addition, timing analysis favored AI‐Rad over manual contouring, indicating the largest time saving (753s per patient) in the Thorax area. AI‐Rad was concluded to be a promising automated contouring solution that generated clinically acceptable contours and achieved time savings, thereby greatly benefiting the radiotherapy process.

Publisher

Wiley

Subject

Radiology, Nuclear Medicine and imaging,Instrumentation,Radiation

Cited by 1 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献