Quantification of the role of lead foil in flattening filter free beam reference dosimetry

Author:

Gao Song1,Nelson Christopher1,Wang Congjun1,Kathriarachchi Vindu1,Choi Michael1,Saxena Rishik1,Kendall Robin1,Balter Peter1

Affiliation:

1. Department of Radiation Physics The University of Texas MD Anderson Cancer Center Houston Texas USA

Abstract

AbstractPurposeTo quantify the potential error in outputs for flattening filter free (FFF) beams associated with use of a lead foil in beam quality determination per the addendum protocol for TG‐51, we examined differences in measurements of the beam quality conversion factor kQ when using or not using lead foil.MethodsTwo FFF beams, a 6 MV FFF and a 10 MV FFF, were calibrated on eight Varian TrueBeams and two Elekta Versa HD linear accelerators (linacs) according to the TG‐51 addendum protocol by using Farmer ionization chambers [TN 30013 (PTW) and SNC600c (Sun Nuclear)] with traceable absorbed dose‐to‐water calibrations. In determining kQ, the percentage depth‐dose at 10 cm [PDD(10)] was measured with 10×10 cm2 field size at 100 cm source‐to‐surface distance (SSD). PDD(10) values were measured either with a 1 mm lead foil positioned in the path of the beam [%dd(10)Pb] or with omission of a lead foil [%dd(10)]. The %dd(10)x values were then calculated and the kQ factors determined by using the empirical fit equation in the TG‐51 addendum for the PTW 30013 chambers. A similar equation was used to calculate kQ for the SNC600c chamber, with the fitting parameters taken from a very recent Monte Carlo study. The differences in kQ factors were compared for with lead foil vs. without lead foil.ResultsDifferences in %dd(10)x with lead foil and with omission of lead foil were 0.9 ± 0.2% for the 6 MV FFF beam and 0.6 ± 0.1% for the 10 MV FFF beam. Differences in kQ values with lead foil and with omission of lead foil were −0.1 ± 0.02% for the 6 MV FFF and −0.1 ± 0.01% for the 10 MV FFF beams.ConclusionWith evaluation of the lead foil role in determination of the kQ factor for FFF beams. Our results suggest that the omission of lead foil introduces approximately 0.1% of error for reference dosimetry of FFF beams on both TrueBeam and Versa platforms.

Publisher

Wiley

Subject

Radiology, Nuclear Medicine and imaging,Instrumentation,Radiation

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3