Energy dependence of x‐ray beam size produced by polycapillary x‐ray optics

Author:

Das Anusheela1ORCID,Heirwegh Chris M.1ORCID,Gao Ning2,Elam William T.3,Wade Lawrence A.1,Clark Benton C.4,Hurowitz Joel A.5,VanBommel Scott J.6,Jones Michael W. M.789,Allwood Abigail C.1

Affiliation:

1. Jet Propulsion Laboratory, California Institute of Technology Pasadena California USA

2. X‐Ray Optical Systems (XOS), Inc. Albany New York USA

3. Applied Physics Laboratory, University of Washington Seattle Washington USA

4. Space Science Institute Boulder Colorado USA

5. Department of Geosciences, Stony Brook University Stony Brook New York USA

6. Department of Earth and Planetary Sciences Washington University in St. Louis St. Louis Missouri USA

7. Central Analytical Research Facility, Queensland University of Technology Brisbane Queensland Australia

8. School of Chemistry and Physics, Queensland University of Technology Brisbane Queensland Australia

9. Center for Planetary Surface Exploration, Queensland University of Technology Brisbane Queensland Australia

Abstract

AbstractIn this work, we studied the x‐ray energy dependence of x‐ray beam diameter focused by polycapillary optics. A quantitative beam diameter–energy relation enables more accurate estimation of the element‐specific interrogation area of a sample using the compositional maps produced by a micro‐XRF system. This improves upon our ability to visualize individual beam‐diameter sized mineral grains and in turn directly benefits Planetary Instrument for X‐ray Lithochemistry (PIXL) analyses of martian soil in addition to benefitting other micro‐focused x‐ray fluorescence (XRF) systems. The spatial distribution of an array of characteristic XRF emission lines was measured by sampling via a knife‐edge approach with small motor stepping of the beam across target edges. Data taken as part of this effort, from the Planetary Flight Model (PFM), were limited to only seven beam energies corresponding to the elements Ni, Cu, Se, Ta, Au, Ti and Ba. Hence, we conducted additional analysis using JPL's lab‐based breadboard (LBB) micro‐XRF system, a system that emulates PIXL's functionality where we measured beam diameter corresponding to 18 elements: Na, Mg, Al, Si, Cl, Ca, Ti, Cr, Mn, Fe, Ni, Cu, Zn, Ga, Ge, Se, Sr and Mo. The experimental results were also compared with Monte Carlo simulations. The beam diameter (y)–energy (x) relation that we obtained for LBB was y = 185.79 exp(−0.078x) whose exponential component was then used to get a more accurate relation for the PFM even with the limited data set: y = 227.53 exp(−0.078x). The difference in the two coefficients for the PFM and LBB stems mainly from the difference in the polycapillary optic design, and this work establishes x‐ray beam diameter versus energy relation quantitatively for both the systems.

Funder

National Aeronautics and Space Administration

Jet Propulsion Laboratory

Publisher

Wiley