Small field output correction factors at 18 MV

Author:

Ringholz Jonas1,Sauer Otto A.1,Wegener Sonja1

Affiliation:

1. Department of Radiation Oncology University Hospital Würzburg Würzburg Germany

Abstract

AbstractBackgroundThe response of various detectors in the radiotherapy energy range has been investigated, especially for 6 and 10 MV energies for small fields, and is summarized in TRS‐483. However, data for accelerator energies above 10 MV are sparse or unavailable for many detectors, especially for the energy of 18 MV. Small variations in field output factors for the commissioning of a treatment planning system can have a high impact on calculation of dose distributions.PurposeMany studies describe an energy dependence of the response for a large number of detectors. We wanted to close the gap for the 18 MV energy regime and determined field output correction factors for different detectors at 18 MV.MethodsAn ELEKTA Versa HD accelerator at 18 MV was used together with a PTW MP3 water phantom at an SSD of 90 cm. The following detectors were examined: PTW Semiflex 31021, PinPoint 3D 31022, diode 60012, diode 60008 and microDiamond 60019, Sun Nuclear EDGE detector, IBA PFD, SFD, Razor Chamber, Razor Nano Chamber and Razor Diode, Standard Imaging Scintillator Exradin W2 1x3, W2 1x1 and Gafchromic EBT3 film. The dose response was determined at a depth of 10 cm for square fields between 0.5 and 10 cm side length. As reference data a composure of radiochromic film data for small fields ( cm) and data of all compatible chambers for larger fields ( cm) was used. The effective field sizes of small fields were determined from profiles obtained on radiochromic film. The obtained field output correction factors obey the rules of the TRS‐483 protocol.ResultsThe W2 1x1 scintillator and the Razor Chamber showed the smallest deviations from the reference curve. The shielded diodes (diode 60008, EDGE detector) showed the highest over‐response at small fields, followed by PFD, microDiamond and the unshielded diodes (diode 60012, SFD). The ionization chambers exhibited the well‐known volume effect, that is, strong under‐response at small fields of up to 9% for the PinPoint 3D, 7% for the Razor Chamber and up to 30% for the Semiflex detector for the smallest studied field size. The small chambers showed a polarity effect in axial orientation, especially the Razor Nano Chamber. Corrections at 18 MV are generally larger than those provided by TRS‐483, continuing the trend of increasing corrections between 6 and 10 MV also at a higher accelerator energy. Only the PinPoint 3D Chamber showed a slightly smaller correction.ConclusionsField output correction factors were determined for square field sizes between 0.5 and 10 cm at 18 MV. Most detectors needed a larger correction than at 6 and 10 MV. Thus, the use of correction factors will improve beam data for 18 MV.

Funder

Deutsche Forschungsgemeinschaft

Publisher

Wiley

Subject

General Medicine

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3