Eigenbin compression for reducing photon‐counting CT data size

Author:

Schmidt Taly Gilat1,Yin Zhye2,Yao Jingwu2,Fan Jiahua2

Affiliation:

1. Department of Biomedical Engineering Marquette University and Medical College of Wisconsin Milwaukee Wisconsin USA

2. GE HealthCare Waukesha Wisconsin USA

Abstract

AbstractBackgroundPhoton‐counting CT (PCCT) systems acquire multiple spectral measurements at high spatial resolution, providing numerous image quality benefits while also increasing the amount of data that must be transferred through the gantry slip ring.PurposeThis study proposes a lossy method to compress photon‐counting CT data using eigenvector analysis, with the goal of providing image quality sufficient for applications that require a rapid initial reconstruction, such as to confirm anatomical coverage, scan quality, and to support automated advanced applications. The eigenbin compression method was experimentally evaluated on a clinical silicon PCCT prototype system.MethodsThe proposed eigenbin method performs principal component analysis (PCA) on a set of PCCT calibration measurements. PCA finds the orthogonal axes or eigenvectors, which capture the maximum variance in the N dimensional photon‐count data space, where N is the number of acquired energy bins. To reduce the dimensionality of the PCCT data, the data are linearly transformed into a lower dimensional space spanned by the M < N eigenvectors with highest eigenvalues (i.e., the vectors that account for most of the information in the data). Only M coefficients are then transferred per measurement, which we term eigenbin values. After transmission, the original N energy‐bin measurements are estimated as a linear combination of the M eigenvectors. Two versions of the eigenbin method were investigated: pixel‐specific and pixel‐general. The pixel‐specific eigenbin method determines eigenvectors for each individual detector pixel, while the more practically realizable pixel‐general eigenbin method finds one set of eigenvectors for the entire detector array. The eigenbin method was experimentally evaluated by scanning a 20 cm diameter Gammex Multienergy phantom with different material inserts on a clinical silicon‐based PCCT prototype. The method was evaluated with the number of eigenbins varied between two and four. In each case, the eigenbins were used to estimate the original 8‐bin data, after which material decomposition was performed. The mean, standard deviation, and contrast‐to‐noise ratio (CNR) of values in the reconstructed basis and virtual monoenergetic images (VMI) were compared for the original 8‐bin data and for the eigenbin data.ResultsThe pixel‐specific eigenbin method reduced photon‐counting CT data size by a factor of four with <5% change in mean values and a small noise penalty (mean change in noise of <12%, maximum change in noise of 20% for basis images). The pixel‐general eigenbin compression method reduced data size by a factor of 2.67 with <5% change in mean values and a less than 10% noise penalty in the basis images (average noise penalty ≤5%). The noise penalty and errors were less for the VMIs than for the basis images, resulting in <5% change in CNR in the VMIs.ConclusionThe eigenbin compression method reduced photon‐counting CT data size by a factor of two to four with less than 5% change in mean values, noise penalty of less than 10%–20%, and change in CNR ranging from 15% decrease to 24% increase. Eigenbin compression reduces the data transfer time and storage space of photon‐counting CT data for applications that require rapid initial reconstructions.

Publisher

Wiley

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3