Randomized double-blind crossover study of alternative stimulator settings in sacral nerve stimulation for faecal incontinence

Author:

Duelund-Jakobsen J12,Dudding T2,Bradshaw E2,Buntzen S1,Lundby L1,Laurberg S1,Vaizey C2

Affiliation:

1. Surgical Research Unit, Department of Surgery P, Aarhus University Hospital, Aarhus, Denmark

2. Physiology Unit, St Mark's Hospital, Harrow, UK

Abstract

Abstract Background Sacral nerve stimulation is an established treatment for faecal incontinence. Nearly half of all patients experience loss of efficacy at some point. Standard reprogramming restores efficacy for some, but not all, patients. This study aimed to determine whether alternative stimulator settings would increase treatment efficacy. Methods Patients with sustained loss of efficacy were recruited from two international specialist centres. A number of alternative stimulation parameters were tested using a double-blind randomized crossover study design. Stimulation settings tested were pulse frequencies of 6·9 and 31 Hz, and pulse widths of 90 and 330 µs, compared with one standard setting of 14 Hz/210 µs. Treatment efficacy was evaluated using a disease-specific quality-of-life score (Fecal Incontinence Quality of Life Scale, FIQLS) and a bowel habit diary completed before randomization, during the study period and after 3 months of follow-up with one preferred setting. Results Fifteen patients were analysed. With one preferred setting, three of four subdomains in the FIQLS improved significantly. The mean(s.d.) total number of incontinence episodes dropped from 11·7(10·8) to 4·8(4·5) per 3 weeks (P = 0·011) and improvements were maintained after 3 months of follow-up. Optimal pacemaker settings were individual, but a trend towards highest patient satisfaction and improved treatment outcome was evident for high-frequency stimulation (31 Hz/210 µs), which was preferred by eight of the 15 patients. Conclusion Patients experiencing loss of efficacy can experience improvement if alternative pacemaker settings are tested. High-frequency stimulation (31 Hz/210 µs) was preferred by more than half of the patients, and improved treatment outcome was sustained at 3 months. Registration number: NCT01254695 (http://www.clinicaltrials.gov).

Publisher

Oxford University Press (OUP)

Subject

Surgery

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3