Free‐breathing abdominal chemical exchange saturation transfer imaging using water presaturation and respiratory gating at 3.0 T

Author:

Chen Zhensen12ORCID,Liu Chuyu3,Wang Yishi4,Guo Rui5,Chen Weibo6,Wang He127,Song Xiaolei3

Affiliation:

1. Institute of Science and Technology for Brain‐Inspired Intelligence Fudan University Shanghai China

2. Key Laboratory of Computational Neuroscience and Brain‐Inspired Intelligence (Fudan University), Ministry of Education Shanghai China

3. Center for Biomedical Imaging Research, Department of Biomedical Engineering School of Medicine Tsinghua University Beijing China

4. Philips Healthcare Beijing China

5. School of Medical Technology Beijing Institute of Technology Beijing China

6. Philips Healthcare Shanghai China

7. Human Phenome Institute Fudan University Shanghai China

Abstract

AbstractFree‐breathing abdominal chemical exchange saturation transfer (CEST) has great potential for clinical application, but its technical implementation remains challenging. This study aimed to propose and evaluate a free‐breathing abdominal CEST sequence. The proposed sequence employed respiratory gating (ResGat) to synchronize the data acquisition with respiratory motion and performed a water presaturation module before the CEST saturation to abolish the influence of respiration‐induced repetition time variation. In vivo experiments were performed to compare different respiratory motion‐control strategies and B0 offset correction methods, and to evaluate the effectiveness and necessity of the quasi‐steady‐state (QUASS) approach for correcting the influence of the water presaturation module on CEST signal. ResGat with a target expiratory phase of 0.5 resulted in a higher structural similarity index and a lower coefficient of variation on consecutively acquired CEST S0 images than breath‐holding (BH) and respiratory triggering (all p < 0.05). B0 maps derived from the abdominal CEST dataset itself were more stable for B0 correction, compared with the separately acquired B0 maps by a dual‐echo time scan and B0 maps derived from the water saturation shift referencing approach. Compared with BH, ResGat yielded more homogeneous magnetization transfer ratio asymmetry maps at 3.5 ppm (standard deviation: 3.96% vs. 3.19%, p = 0.036) and a lower mean squared difference between scan and rescan (27.52‱ vs. 16.82‱, p = 0.004). The QUASS approach could correct the water presaturation‐induced CEST signal change, but its necessity for in vivo scanning needs further verification. The proposed free‐breathing abdominal CEST sequence using ResGat had an acquisition efficiency of approximately four times that using BH. In conclusion, the proposed free‐breathing abdominal CEST sequence using ResGat and water presaturation has a higher acquisition efficiency and image quality than abdominal CEST using BH.

Funder

Fudan University

National Natural Science Foundation of China

Tsinghua University

National Key Research and Development Program of China

Publisher

Wiley

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3