Generation and comparison of 3D dosimetric reference datasets for COMS eye plaque brachytherapy using model‐based dose calculations

Author:

Fletcher Elizabeth M.1,Ballester Facundo23,Beaulieu Luc45,Morrison Hali67,Poher Audran45,Rivard Mark J.8,Sloboda Ron S.910,Vijande Javier2311,Thomson Rowan M.1

Affiliation:

1. Carleton Laboratory for Radiotherapy Physics, Physics Department Carleton University Ottawa Ontario Canada

2. Departamento de Física Atómica, Molecular y Nuclear Universitat de Valencia (UV) Burjassot Spain

3. Unidad Mixta de Investigación en Radiofísica e Instrumentación Nuclear en Medicina (IRIMED) Instituto de Investigación Sanitaria La Fe (IIS‐La Fe)‐Universitat de Valencia (UV) Burjassot Spain

4. Service de physique médicale et Axe oncologie du Centre de recherche du CHU de Québec CHU de Québec Québec Québec Canada

5. Département de Physique, de Génie Physique et d'Optique et Centre de Recherche sur le Cancer Université Laval Québec Québec Canada

6. Department of Oncology and Department of Physics and Astronomy University of Calgary Calgary AB Canada

7. Department of Medical Physics Tom Baker Cancer Centre Calgary Alberta Canada

8. Department of Radiation Oncology Warren Alpert Medical School of Brown University Providence Rhode Island USA

9. Department of Medical Physics Cross Cancer Institute Edmonton Alberta Canada

10. Department of Oncology University of Alberta Edmonton Alberta Canada

11. Instituto de Física Corpuscular IFIC (UV‐CSIC) Burjassot Spain

Abstract

AbstractPurposeA joint Working Group of the American Association of Physicists in Medicine (AAPM), the European Society for Radiotherapy and Oncology (ESTRO), and the Australasian Brachytherapy Group (ABG) was created to aid in the transition from the AAPM TG‐43 dose calculation formalism, the current standard, to model‐based dose calculations. This work establishes the first test cases for low‐energy photon‐emitting brachytherapy using model‐based dose calculation algorithms (MBDCAs).Acquisition and Validation MethodsFive test cases are developed: (1) a single model 6711 125I brachytherapy seed in water, 13 seeds (2) individually and (3) in combination in water, (4) the full Collaborative Ocular Melanoma Study (COMS) 16 mm eye plaque in water, and (5) the full plaque in a realistic eye phantom. Calculations are done with four Monte Carlo (MC) codes and a research version of a commercial treatment planning system (TPS). For all test cases, local agreement of MC codes was within ∼2.5% and global agreement was ∼2% (4% for test case 5). MC agreement was within expected uncertainties. Local agreement of TPS with MC was within 5% for test case 1 and ∼20% for test cases 4 and 5, and global agreement was within 0.4% for test case 1 and 10% for test cases 4 and 5.Data Format and Usage NotesDose distributions for each set of MC and TPS calculations are available online (https://doi.org/10.52519/00005) along with input files and all other information necessary to repeat the calculations.Potential ApplicationsThese data can be used to support commissioning of MBDCAs for low‐energy brachytherapy as recommended by TGs 186 and 221 and AAPM Report 372. This work additionally lays out a sample framework for the development of test cases that can be extended to other applications beyond eye plaque brachytherapy.

Funder

Natural Sciences and Engineering Research Council of Canada

Canada Research Chairs

Ministerio de Ciencia e Innovación

Publisher

Wiley

Subject

General Medicine

Cited by 2 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3