Dynamic range and optimization strategies for radiochromic film calibration using gradient radiation fields

Author:

Pecic Stevan1ORCID,Belca Ivan1,Stojadinovic Strahinja2ORCID,Nidzovic Borko3ORCID,Vicic Milos1,Devic Slobodan45

Affiliation:

1. Faculty of Physics University of Belgrade Belgrade Serbia

2. Department of Radiation Oncology University of Texas Southwestern Medical Center Dallas Texas USA

3. Institute of Oncology and Radiology of Serbia Belgrade Serbia

4. Medical Physics Unit McGill University Montreal Canada

5. Department of Radiation Oncology SMBD Jewish General Hospital Montreal Canada

Abstract

AbstractThis investigation aimed to optimize gradient positioning for radiochromic film calibration to facilitate a uniform distribution of calibration points. The study investigated the influence of various parameters on gradient dose profiles generated by a physical wedge, assessing their impact on the field's dose dynamic range, a scalar quantity representing the span of absorbed doses. Numerical parameterization of the physical wedge profile was used to visualize and quantify the impact of field size, depth, and energy on the dynamic range of dose gradients. This concept enabled the optimization of the gradient positioning and estimation of the necessary number of exposures for the desired calibration dose range. An optimization algorithm based on histogram bin height minimization was developed and presented. The maximum dynamic range was achieved with a 20 20 field size at 5 cm depth. Optimization of wedge gradient positioning yielded the most uniform dose distribution with 7 exposures for the [1,10] Gy range and 8 exposures for the [1,20] Gy range. Film calibration using gradients centered at 1.6, 3, 3.5, and 7 Gy central axis (CAX), obtained through optimized gradient positioning, was showcased. The presented work demonstrates the potential for an improved film calibration process, with efficient material utilization and enhanced dosimetric accuracy for clinical applications. While the method was described for the use of a physical wedge, the methodology can be easily extended to the use of a more convenient dynamic wedge.

Funder

Natural Sciences and Engineering Research Council of Canada

Publisher

Wiley

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3