Evaluating stereotactic accuracy with patient‐specific MRI distortion corrections for frame‐based radiosurgery

Author:

Knill Cory1,Halford Robert1,Sandhu Raminder1,Loughery Brian1,Shamim Sharjil2,Junn Fred1,Lee Kuei1,Almahariq Muayad1,Seymour Zachary1

Affiliation:

1. Department of Radiation Oncology Corewell Health William Beaumont University Hospital Royal Oak Michigan USA

2. William Beaumont School of Medicine Oakland University Rochester Michigan USA

Abstract

AbstractPurposeThis study examines how MRI distortions affect frame‐based SRS treatments and assesses the need for clinical distortion corrections.MethodsThe study included 18 patients with 80 total brain targets treated using frame‐based radiosurgery. Distortion within patients' MRIs were corrected using Cranial Distortion Correction (CDC) software, which utilizes the patient's CT to alter planning MRIs to reduce inherent intra‐cranial distortion. Distortion was evaluated by comparing the original planning target volumes (PTVORIG) to targets contoured on corrected MRIs (PTVCORR). To provide an internal control, targets were also re‐contoured on uncorrected (PTVRECON) MRIs. Additional analysis was done to assess if 1 mm expansions to PTVORIG targets would compensate for patient‐specific distortions. Changes in target volumes, DICE and JACCARD similarity coefficients, minimum PTV dose (Dmin), dose to 95% of the PTV (D95%), and normal tissue receiving 12 Gy (V12Gy), 10 Gy (V10Gy), and 5 Gy (V5Gy) were calculated and evaluated. Student's t‐tests were used to determine if changes in PTVCORR were significantly different than intra‐contouring variability quantified by PTVRECON.ResultsPTVRECON and PTVCORR relative changes in volume were 6.19% ± 10.95% and 1.48% ± 32.92%. PTVRECON and PTVCORR similarity coefficients were 0.90 ± 0.08 and 0.73 ± 0.16 for DICE and 0.82 ± 0.12 and 0.60 ± 0.18 for JACCARD. PTVRECON and PTVCORR changes in Dmin were –0.88% ± 8.77% and −12.9 ± 17.3%. PTVRECON and PTVCORR changes in D95% were −0.34% ± 5.89 and −8.68% ± 13.21%. The 1 mm expanded PTVORIG targets did not entirely cover 14 of the 80 PTVCORR targets. Normal tissue changes (V12Gy, V10Gy, V5Gy) calculated with PTVRECON were (−0.09% ± 7.39%, −0.38% ± 5.67%, −0.08% ± 2.04%) and PTVCORR were (−2.14% ± 7.34%, −1.42% ± 5.45%, −0.61% ± 1.93%). Except for V10Gy, all PTVCORR changes were significantly different (p < 0.05) than PTVRECON.ConclusionMRIs used for SRS target delineation exhibit notable geometric distortions that may compromise optimal dosimetric accuracy. A uniform 1 mm expansion may result in geometric misses; however, the CDC algorithm provides a feasible solution for rectifying distortions, thereby enhancing treatment precision.

Publisher

Wiley

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3