Technical Notes: Robustness of three‐dimensional treatment and imaging isocenter testing using a new gel dosimeter and kilovoltage CBCT

Author:

Oshika Riki12,Tachibana Hidenobu2,Seki Kazuya2,Tachibana Rie3,Moriya Shunsuke4,Sakae Takeji4

Affiliation:

1. Degree Programs in Comprehensive Human Sciences Graduate School of Comprehensive Human Sciences University of Tsukuba Ibaraki Japan

2. Radiation Safety and Quality Assurance division National Cancer Center Hospital East Chiba Japan

3. Triangle Products Co. Ltd. Chiba Japan

4. Faculty of Medicine University of Tsukuba Ibaraki Japan

Abstract

AbstractBackgroundCoincidence of the treatment and imaging isocenter coordinates is required to safely perform small‐margin treatments, such as stereotactic radiosurgery of multiple brain metastases. A comprehensive and direct methodology for verifying concordance of kilovoltage cone‐beam computed tomography (kV‐CBCT) and treatment coordinates using an x‐ray CT‐based polymer gel dosimeter (dGEL) and onboard kV‐CBCT was previously reported. Using this methodology, we tested the ability of a new commercially available x‐ray CT‐based polymer dGEL with a rapid response to provide efficient quality assurance (QA).PurposeThe aim of this study was to evaluate the robustness of the three‐dimensional geometric QA methodology using dGEL.MethodsThe dGEL were commercially manufactured. The prescribed dose for each field was determined by visually identifying the 5, 10, and 20 Gy isodose lines. A linear accelerator was used to irradiate the gels with seven non‐coplanar beams. An in‐house analysis program was used to identify the beam axes and treatment isocenter in kV‐CBCT coordinates by processing the pre‐ and post‐irradiation CBCT images. The impact of the radiation dose on the test reproducibility was examined, and the detectability of an intentional geometric error was assessed.ResultsThe treatment isocenter was within 0.4 mm of the imaging isocenter for all radiation doses. The residual error of the test with the intentional error was within 0.2 mm. The analysis and image quality variations for a single dGEL introduced displacement errors less than 0.3 mm.ConclusionsThe test assessed the coincidence of treatment and kV‐CBCT isocenter coordinates and detected errors with high robustness. Even for a 10 Gy dose, the test yielded results comparable with those obtained using higher radiation doses owing to the rapid response of the dGEL dosimeter.

Publisher

Wiley

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3