A markerless beam's eye view motion monitoring algorithm based on structure conversion and demons registration in medical image with partial data

Author:

Qiu Minmin1,Guan Qi1,Zhong Jiajian1,Huang Taiming1,Luo Ning1,Deng Yongjin1

Affiliation:

1. The First Affiliated Hospital Sun Yat‐Sen University Guangzhou Guangdong Province PR China

Abstract

AbstractPurposeTo propose a markerless beam's eye view (BEV) motion monitoring algorithm, which works with the inferior quality megavolt (MV) images with multi‐leaf collimator (MLC) occlusion‐compatible.MethodsA thorax phantom was used to verify the accuracy of the algorithm. Lung tumor quality assurance (QA) plans were generated for the phantom, and delivered 10 times on the linear accelerator with manually treatment offsets in various directions. The algorithm was used to register 753 electronic portal imaging device (EPID) images with the appropriate digitally reconstructed radiograph (DRR), calculating a registration offset that was compared with the actual offset to determine the monitoring errors. Image similarity measure was used as an independent check. Additionally, patient data of 21 lung tumor treatment plans were gathered. A total of 533 pairs of patient images were acquired for motion monitoring study, to offer quantifiable data of the tumor position change during treatment.ResultsThe monitoring algorithm can process various degrees (10%–80%) of image loss, and performs well when dealing with non‐rigid registration for partial data images. About 86.8% of the monitoring errors are less than 3 mm in the algorithm verification of the phantom study, and about 80% of the errors are under than 2 mm. Normalized Mutual Information (NMI) of phantom images changes from 1.182 ± 0.026 to 1.202 ± 0.027, with p < 0.005, and the Hausdorff‐Distance (HD) changes from 3.506 ± 0.417 mm to 3.466 ± 0.473 mm, with p < 0.005. Translation with a displacement range of ‐6.0 mm to 6.2 mm is the predominant change of the patient target during treatment. NMI of patient images changes from 1.216 ± 0.031 to 1.225 ± 0.031, with p < 0.005, and HD changes from 3.131 ± 0.876 mm to 3.118 ± 0.038 mm, with p < 0.005. The dice index of target before and after registration is 0.264 ± 0.336, indicating the presence of non‐negligible non‐rigid deformation.ConclusionsThe study provides a robust markerless motion monitoring algorithm for multi‐modal, partial data and inferior quality image processing.

Publisher

Wiley

Subject

General Medicine

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3