Multisite MRI Intravoxel Incoherent Motion Repeatability and Reproducibility across 3 T Scanners in a Breast Diffusion Phantom: A BReast Intravoxel Incoherent Motion Multisite (BRIMM) Study

Author:

Basukala Dibash12ORCID,Mikheev Artem1,Sevilimedu Varadan3,Gilani Nima1,Moy Linda1,Pinker Katja3ORCID,Thakur Sunitha B.23ORCID,Sigmund Eric E.1ORCID

Affiliation:

1. Department of Radiology Center for Advanced Imaging and Innovation (CAI2R), Center for Biomedical Imaging, NYU Langone Health New York City New York USA

2. Department of Medical Physics Memorial Sloan Kettering Cancer Center New York City New York USA

3. Department of Radiology Memorial Sloan Kettering Cancer Center New York City New York USA

Abstract

BackgroundMonoexponential apparent diffusion coefficient (ADC) and biexponential intravoxel incoherent motion (IVIM) analysis of diffusion‐weighted imaging is helpful in the characterization of breast tumors. However, repeatability/reproducibility studies across scanners and across sites are scarce.PurposeTo evaluate the repeatability and reproducibility of ADC and IVIM parameters (tissue diffusivity (Dt), perfusion fraction (Fp) and pseudo‐diffusion (Dp)) within and across sites employing MRI scanners from different vendors utilizing 16‐channel breast array coils in a breast diffusion phantom.Study TypePhantom repeatability.PhantomA breast phantom containing tubes of different polyvinylpyrrolidone (PVP) concentrations, water, fat, and sponge flow chambers, together with an MR‐compatible liquid crystal (LC) thermometer.Field Strength/SequenceBipolar gradient twice‐refocused spin echo sequence and monopolar gradient single spin echo sequence at 3 T.AssessmentStudies were performed twice in each of two scanners, located at different sites, on each of 2 days, resulting in four studies per scanner. ADCs of the PVP and water were normalized to the vendor‐provided calibrated values at the temperature indicated by the LC thermometer for repeatability/reproducibility comparisons.Statistical TestsADC and IVIM repeatability and reproducibility within and across sites were estimated via the within‐system coefficient of variation (wCV). Pearson correlation coefficient (r) was also computed between IVIM metrics and flow speed. A P value <0.05 was considered statistically significant.ResultsADC and Dt demonstrated excellent repeatability (<2%; <3%, respectively) and reproducibility (both <5%) at the two sites. Fp and Dp exhibited good repeatability (mean of two sites 3.67% and 5.59%, respectively) and moderate reproducibility (mean of two sites 15.96% and 13.3%, respectively). The mean intersite reproducibility (%) of Fp/Dp/Dt was 50.96/13.68/5.59, respectively. Fp and Dt demonstrated high correlations with flow speed while Dp showed lower correlations. Fp correlations with flow speed were significant at both sites.Data ConclusionIVIM reproducibility results were promising and similar to ADC, particularly for Dt. The results were reproducible within both sites, and a progressive trend toward reproducibility across sites except for Fp.Level of Evidence2Technical EfficacyStage 1

Publisher

Wiley

Subject

Radiology, Nuclear Medicine and imaging

Cited by 2 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3