Whole‐liver flip‐angle shimming at 7 T using parallel‐transmit kT‐point pulses and Fourier phase‐encoded DREAMB1+ mapping

Author:

Runderkamp Bobby A.1ORCID,Roos Thomas23ORCID,van der Zwaag Wietske24ORCID,Strijkers Gustav J.5ORCID,Caan Matthan W. A.5ORCID,Nederveen Aart J.1ORCID

Affiliation:

1. Department of Radiology and Nuclear Medicine Amsterdam University Medical Centers, University of Amsterdam Amsterdam the Netherlands

2. Spinoza Centre for Neuroimaging Royal Netherlands Academy for Arts and Sciences (KNAW) Amsterdam the Netherlands

3. High‐Field Research Group University Medical Center Utrecht Utrecht The Netherlands

4. Computational and Cognitive Neuroscience and Neuroimaging Netherlands Institute for Neuroscience, KNAW Amsterdam the Netherlands

5. Department of Biomedical Engineering and Physics Amsterdam University Medical Centers, University of Amsterdam, Amsterdam Cardiovascular Sciences Amsterdam the Netherlands

Abstract

AbstractPurposeTo obtain homogeneous signal throughout the human liver at 7 T. Flip angle (FA) shimming in 7T whole‐liver imaging was performed through parallel‐transmit kT‐point pulses based on subject‐specific multichannel absolute maps from Fourier phase‐encoded dual refocusing echo acquisition mode (PE‐DREAM).MethodsThe optimal number of Fourier phase‐encoding steps for PE‐DREAM mapping was determined for a 7T eight‐channel parallel‐transmission system. FA shimming experiments were performed in the liver of 7 healthy subjects with varying body mass index. In these subjects, first shimming and Fourier PE‐DREAM mapping were performed. Subsequently, three small‐flip‐angle 3D gradient‐echo scans were acquired, comparing a circularly polarized (CP) mode, a phase shim, and a kT‐point pulse. Resulting homogeneity was assessed and compared with estimated FA maps and distributions.ResultsFourier PE‐DREAM with 13 phase‐encoding steps resulted in a good tradeoff between accuracy and scan time. Lower coefficient of variation values (average [min‐max] across subjects) of the estimated FA in the volume of interest were observed using kT‐points (7.4 [6.6%–8.0%]), compared with phase shimming (18.8 [12.9%–23.4%], p < 0.001) and CP (43.2 [39.4%–47.1%], p < 0.001). kT‐points delivered whole‐liver images with the nominal FA and the highest degree of homogeneity. CP and phase shimming resulted in either inaccurate or imprecise FA distributions. Here, locations having suboptimal FA in the estimated FA maps corresponded to liver areas suffering from inconsistent signal intensity and T1‐weighting in the gradient‐echo scans.ConclusionHomogeneous whole‐liver 3D gradient‐echo acquisitions at 7 T can be obtained with eight‐channel kT‐point pulses calculated based on subject‐specific multichannel absolute Fourier PE‐DREAM maps.