Development of a phantom for assessing the precision of setup in skin mark‐less surface‐guided radiotherapy

Author:

Saito Masahide1ORCID,Ueda Koji1,Nemoto Hikaru1ORCID,Onishi Yoshiko1,Suzuki Hidekazu1,Suzuki Toshihiro2ORCID,Sano Naoki1,Komiyama Takafumi1,Marino Kan1,Onishi Hiroshi1

Affiliation:

1. Department of Radiology University of Yamanashi Yamanashi Japan

2. Kasugai General Rehabilitation Hospital Yamanashi Japan

Abstract

AbstractBackgroundSurface‐guided radiotherapy (SGRT) is adopted by several institutions; however, reports on the phantoms used to assess the precision of the SGRT setup are limited.PurposeThe purpose of this study was to develop a phantom to verify the accuracy of the irradiation position during skin mark‐less SGRT.MethodsAn acrylonitrile butadiene styrene (ABS) plastic cube phantom with a diameter of 150 mm on each side containing a dummy target of 15 mm and two types of body surface‐shaped phantoms (breast/face shape) that could be attached to the cube phantom were fabricated. Films can be inserted on four sides of the cubic phantom (left, right, anterior and posterior), and the center of radiation can be calculated by irradiating the dummy target with orthogonal MV beams. Three types of SGRT using a VOXELAN‐HEV600M (Electronics Research&Development Corporation, Okayama, Japan) were evaluated using this phantom: (i) SGRTCT—a SGRT set‐up based solely on a computed tomography (CT)‐reference image. (ii) SGRTCT + CBCT—a method where cone beam computed tomography (CBCT) matching was performed after SGRTCT. (iii) SGRTScan—a resetup technique using a scan reference image obtained after completing the (ii) step.ResultsBoth the breast and face phantoms were recognized in the SGRT system without problems. SGRTScan ensure precision within 1 mm/1° for breast and face verification, respectively. All SGRT methods showed comparable rotational accuracies with no significant disparities.ConclusionsThe developed phantom was useful for verifying the accuracy of skin mark‐less SGRT position matching. The SGRTScan demonstrated the feasibility of achieving skin‐mark less SGRT with high accuracy, with deviations of less than 1 mm. Additional research is necessary to evaluate the suitability of the developed phantoms for use in various facilities and systems. This phantom could be used for postal surveys in the future.

Publisher

Wiley

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3