Impact of scatter radiation on spectral quantification performance of first‐ and second‐generation dual‐layer spectral computed tomography

Author:

Salazar Edgar12,Liu Leening P.1,Perkins Amy E.3,Halliburton Sandra S.3,Shapira Nadav1,Litt Harold I.1,Noël Peter B.1

Affiliation:

1. Department of Radiology University of Pennsylvania Philadelphia Pennsylvania USA

2. Department of Engineering and Architecture Universidad Privada Boliviana La Paz Bolivia

3. Philips Healthcare Orange Village Ohio USA

Abstract

AbstractObjectiveTo assess the impact of scatter radiation on quantitative performance of first and second‐generation dual‐layer spectral computed tomography (DLCT) systems.MethodA phantom with two iodine inserts (1 and 2 mg/mL) configured to intentionally introduce high scattering conditions was scanned with a first‐ and second‐generation DLCT. Collimation widths (maximum of 4 cm for first generation and 8 cm for second generation) and radiation dose levels were varied. To evaluate the performance of both systems, the mean CT numbers of virtual monoenergetic images (MonoEs) at different energies were calculated and compared to expected values. MonoEs at 50  versus 150 keV were plotted to assess material characterization of both DLCTs. Additionally, iodine concentrations were determined, plotted, and compared against expected values. For each experimental scenario, absolute errors were reported.ResultsAn experimental setup, including a phantom design, was successfully implemented to simulate high scatter radiation imaging conditions. Both CT scanners illustrated high spectral accuracy for small collimation widths (1 and 2 cm). With increased collimation (4 cm), the second‐generation DLCT outperformed the earlier DLCT system. Further, the spectral performance of the second‐generation DLCT at an 8 cm collimation width was comparable to a 4 cm collimation on the first‐generation DLCT. A comparison of the absolute errors between both systems at lower energy MonoEs illustrates that, for the same acquisition parameters, the second‐generation DLCT generated results with decreased errors. Similarly, the maximum error in iodine quantification was less with second‐generation DLCT (0.45  and 0.33 mg/mL for the first and second‐generation DLCT, respectively).ConclusionThe implementation of a two‐dimensional anti‐scatter grid in the second‐generation DLCT improves the spectral quantification performance. In the clinical routine, this improvement may enable additional clinical benefits, for example, in lung imaging.

Funder

National Institutes of Health

Publisher

Wiley

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3