Characterization of the IDENTIFYTM surface scanning system for radiation therapy setup on a closed‐bore linac

Author:

Dekker Janita1ORCID,van het Schip Sander1,Essers Marion1,de Smet Mariska1,Kusters Martijn2,de Kruijf Willy1

Affiliation:

1. Institute Verbeeten Tilburg Netherlands

2. Radboud University Medical Center Nijmegen Netherlands

Abstract

AbstractPurposeIn radiation therapy, surface guidance can be used for patient setup and intra‐fraction motion monitoring. The surface guided radiation therapy (SGRT) system from Varian Medical systems, IDENTIFYTM, consists of three pods, including cameras and a random pattern projector, mounted on the ceiling. The information captured by the cameras is used to make a reconstruction of the surface. The aim of the study was to assess the technical performance of this SGRT system on a closed‐bore linac.MethodsPhantom measurements were performed to assess the accuracy, precision, reproducibility and temporal stability of the system, both in align and in load position. Translations of the phantoms in lateral, longitudinal, and vertical direction, and rotations around three axes (pitch, roll and yaw) were performed with an accurate, in‐house built, positioning stage. Different phantom geometries and different surface colors were used, and various ambient light intensities were tested.ResultsThe accuracy of the IDENTIFYTM system at the closed‐bore linac was 0.07 mm and 0.07 degrees at load position, and 0.06 mm and 0.01 degrees at align position for the white head phantom. The precision was 0.02 mm and 0.02 degrees in load position, and 0.01 mm and 0.02 degrees in align position. The accuracy for the Penta‐Guide phantom was comparable to the white head phantom, with 0.06 mm and 0.01 degrees in align position. The system was slightly less accurate for translations of the CIRS lung phantom in align position (0.20 mm, 0.05 degrees). Reproducibility measurements showed a variation of 0.02 mm in load position. Regarding the temporal stability, the maximum drift over 30 min was 0.33 mm and 0.02 degrees in load position. No effect of ambient light level on the accuracy of the IDENTIFYTM system was observed. Regarding different surface colors, the accuracy of the system for a black phantom was slightly worse compared to a white surface, but not clinical relevant.ConclusionThe IDENTIFYTM system can adequately be used for motion monitoring on a closed‐bore linac with submillimeter accuracy. The results of the performed measurements meet the clinical requirements described in the guidelines of the AAPM and the ESTRO.

Publisher

Wiley

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3