The impact of automatic tube current modulation related settings of a modern GE CT scanner on image quality and patient dose; details do matter

Author:

Tsalafoutas Ioannis A.1ORCID,AlKhazzam Shady1,Kharita Mohammed Hassan1

Affiliation:

1. Medical Physics Section Occupational Health and Safety Department Hamad Medical Corporation Doha Qatar

Abstract

AbstractPurposeTo investigate the operation principles of the automatic tube current modulation (ATCM) of a modern GE healthcare CT scanner, and the impact of related settings on image quality and patient dose.Material & MethodsA dedicated phantom (Mercury 4.0) was scanned using two of the most frequently used clinical scanning protocols (chest and abdomen‐pelvis). The preset protocol settings were used as starting points (reference conditions). Scan direction, scan mode (helical vs. axial), total beam width, tube potential (kVp), and ATCM settings were then modified individually to understand their impact on radiation dose and image quality. Regarding the ATCM settings, the SmartmA minimum and maximum mA limits, and the noise index (NI) values were varied. As surrogates of patient dose, the CTDIvol and DLP values of each scan were used. As surrogates of image quality were used the image noise and the detectability index (d’) of five different materials (air, solid water, polystyrene, iodine, and bone) embedded in the Mercury phantom calculated with the ImQuest software.ResultsThe scanning direction did not have any effect on ATCM curves, unlike what has been observed in CT scanners from other manufacturers. Total beam width does matter, however, the SmartmA limit settings and kVp selection had the greatest impact on image quality and dose. It was seen that improper minimum mA limit settings practically invalidated the ATCM operation. In contrast, when full modulation was allowed without restrictions, noise standard deviation, and detectability index became much more consistent across the wide range of phantom diameters. For lower kVp settings an impressive dose reduction was observed that requires further investigation.ConclusionSmartmA is a tool that if not properly used may increase the patient doses considerably. Therefore, its settings should be carefully adjusted for each preset different clinical protocol.

Publisher

Wiley

Cited by 1 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3