Abstract
AbstractIon mobility–mass spectrometry (IM–MS) is an ever‐evolving tool to separate ions in the gas phase according to electrophoretic mobility with subsequent mass determination. CE is rarely coupled to IM–MS, possibly due to similar separation mechanisms based on electrophoretic mobility. Here, we investigate the orthogonality of CE and ion mobility (IM) by analyzing a complex peptide mixture (tryptic digest of HeLa proteins) with trapped ion mobility mass spectrometry (TIMS–MS). Using the nanoCEasy interface, excellent sensitivity was achieved by identifying thousands of peptides and achieving a peak capacity of 7500 (CE: 203–323 in a 150 cm long capillary, IM: 27–31). Plotting CE versus mass and CE versus (inverse) mobility, a clear grouping in curved striped patterns is observed according to the charge‐to‐size and mass‐to‐charge ratios. The peptide charge in the acidic background electrolyte can be estimated from the number of basic amino acids, with a few exceptions where neighboring effects reduce the positive charge. A surprisingly high orthogonality of CE and IM is observed, which is obviously caused by solvation effects leading to different charges and sizes in the liquid phase compared to the gas phase. A high orthogonality of CE and ion mobility is expected to be observed for other peptide samples as well as other substance classes, making CE–IM–MS a promising tool for various applications.
Subject
Clinical Biochemistry,Biochemistry,Analytical Chemistry
Cited by
1 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献
1. Trendbericht Analytische Chemie 2024;Nachrichten aus der Chemie;2024-03-29