Multibody dynamic modeling of the behavior of flexible instruments used in cervical cancer brachytherapy

Author:

Straathof Robin12,Meijaard Jaap P.3,van Vliet‐Pérez Sharline M.12,Kolkman‐Deurloo Inger‐Karine K.2,Nout Remi A.2,Heijmen Ben J. M.2,Wauben Linda S. G. L.1,Dankelman Jenny1,van de Berg Nick J.14

Affiliation:

1. Department of BioMechanical Engineering Delft University of Technology Delft the Netherlands

2. Department of Radiotherapy Erasmus MC Cancer Institute University Medical Center Rotterdam Rotterdam the Netherlands

3. Department of Precision and Microsystems Engineering Delft University of Technology Delft the Netherlands

4. Department of Gynecological Oncology Erasmus MC Cancer Institute University Medical Center Rotterdam Rotterdam the Netherlands

Abstract

AbstractBackgroundThe steep radiation dose gradients in cervical cancer brachytherapy (BT) necessitate a thorough understanding of the behavior of afterloader source cables or needles in the curved channels of (patient‐tailored) applicators.PurposeThe purpose of this study is to develop and validate computer models to simulate: (1) BT source positions, and (2) insertion forces of needles in curved applicator channels. The methodology presented can be used to improve the knowledge of instrument behavior in current applicators and aid the development of novel (3D‐printed) BT applicators.MethodsFor the computer models, BT instruments were discretized in finite elements. Simulations were performed in SPACAR by formulating nodal contact force and motion input models and specifying the instruments’ kinematic and dynamic properties. To evaluate the source cable model, simulated source paths in ring applicators were compared with manufacturer‐measured source paths. The impact of discrepancies on the dosimetry was estimated for standard plans. To validate needle models, simulated needle insertion forces in curved channels with varying curvature, torsion, and clearance, were compared with force measurements in dedicated 3D‐printed templates.ResultsComparison of simulated with manufacturer‐measured source positions showed 0.5–1.2 mm median and <2.0 mm maximum differences, in all but one applicator geometry. The resulting maximum relative dose differences at the lateral surface and at 5 mm depth were 5.5% and 4.7%, respectively. Simulated insertion forces for BT needles in curved channels accurately resembled the forces experimentally obtained by including experimental uncertainties in the simulation.ConclusionThe models developed can accurately predict source positions and insertion forces in BT applicators. Insights from these models can aid novel applicator design with improved motion and force transmission of BT instruments, and contribute to the estimation of overall treatment precision. The methodology presented can be extended to study other applicator geometries, flexible instruments, and afterloading systems.

Funder

KWF Kankerbestrijding

Publisher

Wiley

Subject

General Medicine

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3