Pulsed wave Doppler ultrasound: Accuracy, variability, and impact of acquisition parameters on flow measurements

Author:

Russ Megan K.12,Lafata Nicole M.12,Robertson Scott H.12,Samei Ehsan1234

Affiliation:

1. Clinical Imaging Physics Group Duke University Health System Durham North Carolina USA

2. Department of Radiology Duke University Durham North Carolina USA

3. Center for Virtual Imaging Trials Ravin Advanced Imaging Laboratories Department of Radiology Duke University Durham North Carolina USA

4. Department of Electrical and Computer Engineering Duke University Durham North Carolina USA

Abstract

AbstractBackgroundPulsed wave Doppler ultrasound is a useful modality for assessing vascular health as it quantifies blood flow characteristics. To facilitate accurate diagnosis, accuracy and consistency of this modality should be assessed through Doppler quality assurance (QA).PurposeThe purpose of this study was to characterize the accuracy, reproducibility, and inter‐scanner variability of ultrasound flow velocity measurements via a flow phantom, with a focus on the effect of systematic acquisition parameters on measured flow velocity accuracy.MethodsUsing a manufacturer‐calibrated flow phantom, pulsed wave measurements were acquired on five clinical systems (iU22, Philips) with three models of transducers, including both linear and curvilinear models. The peak and mean flow velocities were estimated by vendor‐supplied spectral analysis tools. To investigate intra‐ and inter‐scanner variability, measurements were repeated using each scanner‐transducer pair under a standardized set of conditions. Inter‐scanner variability was assessed using ANOVA. Flow velocity accuracy was investigated by mean absolute percentage error. The impacts of receive gain, measurement depth, and beam steering on measured flow velocity accuracy were examined by varying each parameter over its available range and comparing to the ground truth flow velocity.ResultsInter‐scanner variability was statistically significant for peak flow measurements made using both linear and curvilinear transducers, though absolute differences in measured velocity were small. Inter‐scanner variability was not statistically significant for mean flow velocity. Receive gain, measurement depth, and beam steering were all found to impact the accuracy of measured flow characteristics for linear transducers. Accuracy of the flow measurements made with the curvilinear transducer demonstrated high consistency to changes in receive gain at a constant depth, though were impacted by increasing the measurement depth.ConclusionsCarefully and consistently selected acquisition and set‐up parameters are essential in order to establish a reliable and meaningful QA program.

Publisher

Wiley

Subject

General Medicine

Reference18 articles.

1. America RSoN.Ultrasound—Vascular 2022. Updated July 30 2021.https://www.radiologyinfo.org/en/info/vascularus

2. MitchellE Noninvasive diagnosis of upper and lower extremity arterial disease UpToDate2022. Updated 9/1/2022.https://www.uptodate.com/contents/noninvasive‐diagnosis‐of‐upper‐and‐lower‐extremity‐arterial‐disease

3. Carotid Artery Stenosis: Gray-Scale and Doppler US Diagnosis—Society of Radiologists in Ultrasound Consensus Conference

4. Overestimation of velocity and frequency values by multielement linear array Dopplers;Daigle R;J Vasc Technol,1990

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3