Proof‐of‐concept for a thin conical X‐ray target optimized for intensity and directionality for use in a carbon nanotube‐based compact X‐ray tube

Author:

Insley Ben1,Bartkoski Dirk2,Balter Peter1,Prajapati Surendra1,Tailor Ramesh1,Salehpour Mohammad1,Jaffray David3

Affiliation:

1. Department of Radiation Physics The University of Texas MD Anderson Cancer Center Houston Texas USA

2. Empyrean Medical Systems, Inc. 950 Peninsula Corp Cir Boca Raton USA

3. Division of Office of the Sr. VP & Chief Technology and Digital Officer The University of Texas MD Anderson Cancer Center Houston Texas USA

Abstract

AbstractBackgroundCarbon nanotube‐based cold cathode technology has revolutionized the miniaturization of X‐ray tubes. However, current applications of these devices required optimization for large, uniform fields with low intensity.PurposeThis work investigated the feasibility and radiological characteristics of a novel conical X‐ray target optimized for high intensity and high directionality to be used in a compact X‐ray tube.MethodsThe proposed device uses an ultrathin, conical tungsten‐diamond target that exhibits significant heat loading while maintaining a small focal spot size and promoting forward‐directedness of the X‐ray field through preferential attenuation of oblique‐angled photons. The electrostatic and thermal properties of the theoretical tube were calculated and analyzed using COMSOL Multiphysics software. The production, transport, and calculation of radiological properties associated with the resultant X‐ray field were performed using the Geant4 toolkit via its wrapper, TOPAS.ResultsHeat transfer analysis of this X‐ray tube demonstrated the feasibility of a 200‐kV electron beam bombarding the proposed target at a maximum current of 100 mA using a 1‐ms symmetric duty cycle. The cathode of the X‐ray tube was designed to be segmented into nine switchable electrical segments for modulation of the focal spot size from 0.4‐ to 10.8‐mm. After importing the COMSOL‐derived electron beam into TOPAS for X‐ray production simulations, radiological analysis of the resultant field demonstrated high levels of intrinsic beam collimation while maintaining high intensity. A maximum dose rate of 17,887 cGy/min was calculated for 1‐mm depth in water at 7‐cm distance.ConclusionsThe proposed X‐ray tube design can create highly directional X‐ray fields with superior fluence compared to that of current commercial X‐ray tubes of comparable size.

Publisher

Wiley

Subject

General Medicine

Reference40 articles.

1. Field Emission Cathodes to Form an Electron Beam Prepared from Carbon Nanotube Suspensions

2. High-Performance Cold Cathode X-ray Tubes Using a Carbon Nanotube Field Electron Emitter

3. Reyes‐MenaA JensenC BardE et al.Miniature X‐ray tubes utilizing carbon‐nanotube‐based cold cathodes2005;48:6

4. PesceM WangX RowlandP Using carbon nano‐tube field emitters to miniaturize X‐ray tubes. n.d.; 5.