Multiecho–based fat artifact correction for CEST MRI at 7 T

Author:

Tkotz Katharina1ORCID,Liebert Andrzej1ORCID,Gast Lena V.1ORCID,Zeiger Paula1,Uder Michael1,Zaiss Moritz23ORCID,Nagel Armin M.14ORCID

Affiliation:

1. Institute of Radiology, University Hospital Erlangen, Friedrich‐Alexander‐Universität Erlangen‐Nürnberg (FAU) Erlangen Germany

2. Department of Neuroradiology University Hospital Erlangen, Friedrich‐Alexander‐Universität Erlangen‐Nürnberg (FAU) Erlangen Germany

3. Magnetic Resonance Center, Max Planck Institute for Biological Cybernetics Tübingen Germany

4. Medical Physics in Radiology, German Cancer Research Center (DKFZ) Heidelberg Germany

Abstract

AbstractPurposeCEST MRI is influenced by fat signal, which can reduce the apparent CEST contrast or lead to pseudo‐CEST effects. Our goal was to develop a fat artifact correction based on multi‐echo fat–water separation that functions stably for 7 T knee MRI data.MethodsOur proposed algorithm utilizes the full complex data and a phase demodulation with an off‐resonance map estimation based on the Z‐spectra prior to fat–water separation to achieve stable fat artifact correction. Our method was validated and compared to multi‐echo–based methods originally proposed for 3 T by Bloch–McConnell simulations and phantom measurements. Moreover, the method was applied to in vivo 7 T knee MRI examinations and compared to Gaussian fat saturation and a published single‐echo Z‐spectrum–based fat artifact correction method.ResultsPhase demodulation prior to fat–water separation reduced the occurrence of fat–water swaps. Utilizing the complex signal data led to more stable correction results than working with magnitude data, as was proposed for 3 T. Our approach reduced pseudo‐nuclear Overhauser effects compared to the other correction methods. Thus, the mean asymmetry contrast at 3.5 ppm in cartilage over five volunteers increased from −9.2% (uncorrected) and −10.6% (Z‐spectrum–based) to −1.5%. Results showed higher spatial stability than with the fat saturation pulse.ConclusionOur work demonstrates the feasibility of multi‐echo–based fat–water separation with an adaptive fat model for fat artifact correction for CEST MRI at 7 T. Our approach provided better fat artifact correction throughout the entire spectrum and image than the fat saturation pulse or Z‐spectrum–based correction method for both phantom and knee imaging results.

Publisher

Wiley

Subject

Radiology, Nuclear Medicine and imaging

Cited by 1 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3