The Accuracy and Precision of the Continuously Stored Data from Flash Glucose Monitoring System in Type 2 Diabetes Patients during Standard Meal Tolerance Test

Author:

Yan Rengna1,Li Huiqin1ORCID,Kong Xiaocen1ORCID,Zhai Xiaofang1,Chen Maoyuan1,Sun Yixuan1,Ye Lei2ORCID,Su Xiaofei1ORCID,Ma Jianhua1ORCID

Affiliation:

1. Department of Endocrinology, Nanjing First Hospital, Nanjing Medical University, Nanjing 210012, China

2. National Heart Research Institute Singapore, National Heart Centre Singapore, Singapore

Abstract

Background. The purpose of this study was to investigate the accuracy of the continuously stored data from the Abbott FreeStyle Libre flash glucose monitoring (FGM) system in Chinese diabetes patients during standard meal tests when glucose concentrations were rapidly changing. Subjects and Methods. Interstitial glucose levels were monitored for 14 days in 26 insulin-treated patients with type 2 diabetes using the FGM system. Standard meal tests were conducted to induce large glucose swings. Venous blood glucose (VBG) was tested at 0, 30, 60, and 120 min after standard meal tests in one middle day of the first and second weeks, respectively. The corresponding sensor glucose values were obtained from interpolating continuously stored data points. Assessment of accuracy was according to recent consensus recommendations with median absolute relative difference (MARD) and Clarke and Parkes error grid analysis (CEG and PEG). Results. Among 208 paired sensor-reference values, 100% were falling within zones A and B of the Clarke and Parkes error grid analysis. The overall MARD was 10.7% (SD, 7.8%). Weighted least squares regression analysis resulted in high agreement between the FGM sensor glucose and VBG readings. The overall MTT results showed that FGM was lower than actual VBG, with MAD of 22.1 mg/dL (1.2 mmol/L). At VBG rates of change of -1 to 0, 0 to 1, 1 to 2, and 2 to 3 mg/dl/min, MARD results were 11.4% (SD, 8.7%), 9.4% (SD, 6.5%), 9.9% (SD, 7.5%), and 9.5% (SD, 7.7%). At rapidly changing VBG concentrations (>3 mg/dl/min), MARD increased to 19.0%, which was significantly higher than slow changing BG groups. Conclusions. Continuously stored interstitial glucose measurements with the FGM system were found to be acceptable to evaluate VBG in terms of clinical decision during standard meal tests. The continuously stored data from the FGM system appeared to underestimate venous glucose and performed less well during rapid glucose changes.

Funder

National Key R&D Program of China

Publisher

Hindawi Limited

Subject

Endocrine and Autonomic Systems,Endocrinology,Endocrinology, Diabetes and Metabolism

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3