Abstract

Background. Robotic locomotion rehabilitation systems have been used for gait training in patients who have had a stroke. Most commercialized systems allow patients to perform simple exercises such as balancing or level walking, but an additional function such as stair-walk training is required to provide a wide range of recovery cycle rehabilitation. In this study, we analyzed stair-gait patterns and applied the result to a robotic rehabilitation system that can provide a vertical motion of footplates. Methods. To obtain applicable data for the robotic system with vertically movable footplates, stair-walk action was measured using an optical marker-based motion capture system. The spatial position data of joints during stair walking was obtained from six healthy adults who participated in the experiment. The measured marker data were converted into joint kinematic data by using an algorithm that included resampling and normalization. The spatial position data are represented as angular trajectories and the relative displacement of each joint on the anatomical sagittal plane and movements of hip joints on the anatomical transverse plane. Results. The average range of motion (ROM) of each joint was estimated as (6.75°,48.69°) at the hip, 8.20°,93.78° at the knee, and 17.78°,11.75° at the ankle during ascent and as 6.41°,31.67° at the hip, 7.38°,91.93° at the knee, and 24.89°,24.18° at the ankle during descent. Additionally, we attempted to create a more natural stair-gait pattern by analyzing the movement of the hip on the anatomical transverse plane. The hip movements were estimated to within ±1.57cm and ±2.00cm for hip translation and to within ±2.52° and ±2.70° for hip rotation during stair ascent and stair descent, respectively. Conclusions. Based on the results, standard patterns of stair ascent and stair descent were derived and applied to a lower-limb rehabilitation robot with vertically movable footplates. The relative trajectory from the experiment ascertained that the function of stair walking in the robotic system properly worked within a normal ROM.

Funder

Ministry of Health and Welfare

Publisher

Hindawi Limited

Subject

Biomedical Engineering,Bioengineering,Medicine (miscellaneous),Biotechnology

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3