Affiliation:
1. Department of Ophthalmology, Toho University Omori Medical Center, 6-11-1 Omori-nishi, Ota-ku, Tokyo 143-8541, Japan
2. Department of Neonatology, Toho University Omori Medical Center, 6-11-1 Omori-nishi, Ota-ku, Tokyo 143-8541, Japan
Abstract
Measuring the ocular blood flow in neonates may clarify the relationships between eye diseases and ocular circulation abnormalities. However, no method for noninvasively measuring ocular circulation in neonates is established. We used laser speckle flowgraphy (LSFG) modified for neonates to measure their ocular circulation and investigated whether this method is reproducible. During their normal sleep, we studied 16 subjects (adjusted age of 34–48 weeks) whose blood flow could be measured three consecutive times. While the subjects slept in the supine position, three mean blur rate (MBR) values of the optic nerve head (ONH) were obtained: the MBR-A (mean of all values), MBR-V (vessel mean), and MBR-T (tissue mean), and nine blood flow pulse waveform parameters in the ONH were examined. We analyzed the coefficient of variation (COV) and the intraclass correlation coefficient (ICC) for each parameter. The COVs of the MBR values were all ≤10%. The ICCs of the MBR values were all >0.8. Good COVs were observed for the blowout score, blowout time, rising rate, falling rate, and acceleration time index. Although the measurement of ocular circulation in the neonates was difficult, our results exhibited reproducibility, suggesting that this method could be used in clinical research.
Subject
General Immunology and Microbiology,General Biochemistry, Genetics and Molecular Biology,General Medicine
Cited by
10 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献