The Canine Erythrocyte Sedimentation Rate (ESR): Evaluation of a Point-of-Care Testing Device (MINIPET DIESSE)

Author:

Militello Carlo1,Pasquini Anna1,Medina Valentin Anyela Andrea1,Simčič Petra1,De Feo Giulia1,Lubas George1ORCID

Affiliation:

1. Department of Veterinary Science, University of Pisa, Pisa, Italy

Abstract

The erythrocyte sedimentation rate (ESR) in canine medicine has been replaced by the evaluation of a few sensitive markers of the acute-phase proteins. The aim of the study was to evaluate the ESR using a point-of-care (MINIPET, DIESSE Diagnostica Senese S.p.A.) device (ESR-MP) and to compare the results with the gold standard Westergren method (ESR-W) in dogs. One hundred and nineteen K3-EDTA blood samples for complete blood count were randomly selected and assayed for ESR. The reference interval (RI) was established using the percentile method. The coefficient of variation (CV) in intra-assay and interassay precision of ESR-MP was calculated. The analytical sensitivity (Se), specificity (Sp), positive predictive values (PPVs), and negative predictive values (NPVs) were calculated. Agreement between ESR-MP and ESR-W was assessed with Pearson correlation coefficient (r), Cohen concordance test (K), Passing-Bablok regression, and Bland–Altman plots. Ten canine samples (8.4%) were ruled out because of flag-error by the MINIPET instrument (4.2%) or because they showed the diphasic pattern in ESR-W (4.2%). The canine RI of ESR-MP was 0–10 mm/h. Precision was excellent in intra-assay (CV = 0.02) and interassay (CV = 0.32). The analytical characteristics of ESR-MP in nonanemic samples were as follows: Se = 0.82, Sp = 0.95, PPV = 0.82, and NPV = 0.95. The accuracy of ESR-MP was better than ESR-W in nonanemic samples (r = 0.87; K = 0.77) and lower in anemic subjects (Hct <37%) (r = 0.76; K = 0.69). The Passing-Bablok regression showed the presence of systematic error and the absence of proportional error only in nonanemic blood samples. The Bland–Altman plots gave negative average values due to the difference in RIs and an agreement in both ESRs. The ESR-MP results can be obtained with the same K3-EDTA tubes used for the blood count, in shortcut time, and at reduced costs using the MINIPET device. These investigations highlight that ESR-MP could be useful in canine clinical settings.

Funder

Università di Pisa

Publisher

Hindawi Limited

Subject

General Veterinary

Cited by 4 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3