Affiliation:
1. Department of Mechanics, Materials and Biomedical Engineering, Faculty of Mechanical Engineering, Wroclaw University of Science and Technology, Wroclaw, Poland
2. Department of Neurosurgery, Wroclaw Medical University, Wroclaw, Poland
Abstract
The aim of this experimental study was to analyze the impact of applying different configurations of the transpedicular fixation system on selected mechanical parameters of the thoracolumbar spine under conditions of its instability (after simulated fracture). Five study groups were tested: physiological, with compression fracture of the vertebra, with two-segment fixation, with three-segment fixation, and with four-segment fixation. Each of the analyzed study groups was subjected to axial compression, flexion, and extension. Based on the conducted experimental tests, the mechanical parameters, i.e., stiffness coefficient and dissipation energy, were determined for all groups under consideration. The stiffness value of two-segment fixation is significantly lower than the physiological value (during flexion and extension). The use of long-segment fixation considered in two configurations (three- and four-segment fixation) may result in excessive stiffness of the system due to the high stiffness values achieved (approx. 25–30% higher than the physiological values in the case of compression and on average 60% higher in the case of flexion). The use of long-segment fixator design shows better results than short-segment fixation. Considering both biomechanical and clinical aspects, three-segment fixation seems to be a compromise solution as it saves the patient from more extensive stiffening of the spinal motion segments.
Subject
Biomedical Engineering,Bioengineering,Medicine (miscellaneous),Biotechnology
Cited by
3 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献