Comparison of IOP Measurement by Goldmann Applanation Tonometer, ICare Rebound Tonometer, and Tono-Pen in Keratoconus Patients after MyoRing Implantation

Author:

Rateb Mahmoud12ORCID,Abdel-Radi Mahmoud12ORCID,Eldaly Zeiad12,Elmohamady Mohamed Nagy34ORCID,Noor El Din Asaad25

Affiliation:

1. Department of Ophthalmology, Faculty of Medicine, Assiut University, Assiut, Egypt

2. Tiba Eye Center, Assiut, Egypt

3. Department of Ophthalmology, Benha University, Benha, Egypt

4. Masa Eye Center, Benha, Egypt

5. Department of Ophthalmology, Al Azhar University, Cairo, Egypt

Abstract

Purpose. To evaluate the different IOP readings by Goldmann applanation tonometer (GAT), ICare rebound tonometer, and Tono-Pen in keratoconus patients after MyoRing implantation. To assess the influence of central corneal thickness (CCT) and thinnest corneal location (TCL) on IOP measurements by different tonometers. Setting. Prospective observational study was conducted in two private centers in Egypt from February 2015 to November 2016. Methods. Seventeen eyes of 10 patients suffering from keratoconus and who underwent MyoRing implantation were recruited. All subjects underwent GAT, ICare, and Tono-Pen IOP measurements in random order. Central corneal thickness and thinnest corneal location were assessed by Pentacam. Difference in mean in IOP readings was assessed by T-test. Correlation between each pair of devices was evaluated by Pearson correlation coefficient. The Bland–Altman analysis was used to assess intertonometer agreement. Results. Seventeen eyes (10 patients) were evaluated. The mean IOP reading was 13.9 ± 3.68, 12.41 ± 2.87, and 14.29 ± 1.31 mmHg in GAT, ICare, and Tono-Pen group, respectively. There was a significant difference between IOP readings by GAT/ICare and Tono-Pen/ICare (p value: 0.032 and 0.002, respectively) with no significant difference between GAT/Tono-Pen (p value: 0.554). Mean difference in IOP measurements between GAT/ICare was 1.49 ± 2.61 mmHg, Tono-Pen/ICare was 1.89 ± 2.15 mmHg, and GAT/Tono-Pen was −0.39 ± 2.59 mmHg. There was no significant correlation between the difference in IOP readings among any pair of devices and CCC or TCL. The Bland–Altman analysis showed a reasonable agreement between any pair of tonometers.

Publisher

Hindawi Limited

Subject

Ophthalmology

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3