Abstract
In this work, we investigated different short molecule polymer coatings in piezoelectric ceramic-polymer composites with low fibre volume contents. Modifying the interphase between the piezoelectric PZT (lead zirconate titanate) fibre and the epoxy matrix thus enhances the electromechanical coupling factor for 1–3 ultrasound transducers with low fibre contents. It is known that the electromechanical coupling factor can be increased by precoating a ceramic fibre with a soft interlayer polymer [1-1-3]. In this paper, we investigate the so-called 1-1-1-3 composites composed of a ferroelectric ceramic fibre (core), a soft polymer layer (e.g., fatty acids, amides, waxes, or oils), an epoxy resin shell, and an epoxy resin matrix. Some soft polymer layers allowed the free movement of the ferroelectric fibres reducing blocking or clamping by the inactive polymeric matrix, resulting in higher electromechanical coupling factors (kt) for composites with low fibre volume contents. Using an oil-based interlayer, the dielectric constant can be significantly increased. The lowest fibre push-out stress could be achieved with the paraffin interlayer; however, no correlation with the coupling factor could be observed.