Comparison of Anterior Ocular Biometric Measurements Using Swept-Source and Time-Domain Optical Coherence Tomography

Author:

Chen Sisi1ORCID,Gao Rongrong1,McAlinden Colm12,Ye Junming1,Wang Yiran1,Chen Min1,Huang Jinhai1,Sun Yong3ORCID,Yu A-Yong1ORCID

Affiliation:

1. School of Ophthalmology and Optometry and Eye Hospital, Wenzhou Medical University, Wenzhou, Zhejiang, China

2. Department of Ophthalmology, Singleton Hospital, Swansea Bay University Health Board, Port Talbot, UK

3. Shenzhen Hospital of Integrated Traditional and Western Medicine, Shenzhen, China

Abstract

Purpose. To compare central corneal thickness (CCT), aqueous depth (AQD), and anterior chamber depth (ACD) measurements using the swept-source (CASIA SS-1000, Tomey, Japan) and time-domain (Visante, Carl Zeiss Meditec, USA) anterior segment optical coherence tomographers (OCT) in normal eyes. Methods. Sixty-eight eyes of 68 subjects were included. Three consecutive scans of each subject were obtained using both devices in a random order by one experienced operator. Standard deviation (Sw), coefficient of repeatability (CoR), coefficients of variation (CoV), and intraclass correlation coefficients (ICC) were used to evaluate the intraoperator repeatability. Agreement was assessed using the Bland–Altman plots and 95% limits of agreement (LoA). Results. All measurements of the swept-source OCT (SS-OCT) and time-domain OCT (TD-OCT) showed high repeatability with low CoR (CCT: 2.34 μm and 6.16 μm; AQD: 0.05 mm and 0.09 mm; ACD: 0.06 mm and 0.09 mm), low CoV (CCT: 0.16% and 0.42%; AQD: 0.61% and 0.97%; ACD: 0.53% and 0.83%), and high ICC (>0.98). The mean CCT with SS-OCT was slightly thicker than the results with TD-OCT (difference = 4.55 ± 2.62 μm, P<0.001). There was no statistically significant difference in AQD or ACD measurements between the two devices (0.01 ± 0.05 mm, P=0.111; 0.02 ± 0.05 mm, P=0.022, respectively). The 95% LoA between the SS-OCT and TD-OCT were −0.59 to 9.69 μm for CCT, −0.10 to 0.12 mm for AQD, and −0.09 to 0.12 mm for ACD. Conclusions. High levels of repeatability and agreement were found between the two devices for all three parameters, suggesting interchangeability. SS-OCT demonstrated superior repeatability compared with TD-OCT.

Funder

Foundation of Wenzhou City Science & Technology Bureau

Publisher

Hindawi Limited

Subject

Ophthalmology

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3