One-Year Clinical Outcome of Inspiron Stent in All-Comers Population (Analysis from 790 Consecutive Patients)

Author:

Falcão Felipe123ORCID,Cantarelli Fabiano12ORCID,Cantarelli Rodrigo12ORCID,Mota Flávio12ORCID,Navarro Manuela12ORCID,Mota Henrique1ORCID,Santos Martinelly12ORCID,Cruz Daniel2ORCID,Sansônio André1ORCID,Parente Marcelo1ORCID,Oliveira Flávio12ORCID

Affiliation:

1. Hospital Dom Helder Câmara, Cabo de Santo Agostinho, Brazil

2. Unidade de Cardiologia Invasiva (UCI)-Hospital Memorial São José, Rede D’Or São Luiz, Recife, Brazil

3. Department of Internal Medicine, University of Pernambuco, Garanhuns, Brazil

Abstract

Aims. The goal of this study was to evaluate the performance of the InspironTM coronary stent (Scitech Medical™, Goiás, Brazil). The InspironTM sirolimus-eluting stent uses an ultrathin L-605 cobalt-chromium alloy with a 75 μm strut thickness platform coated with an abluminal biodegradable polymer. The polymer is eliminated from the body through the tricarboxylic acid cycle in 6–9 months, releasing 80% of the drug within 30 days after its deployment. Methods. It was a prospective, single-center registry. To represent clinical practice, all patients undergoing percutaneous coronary intervention were included in this registry. There were no exclusion criteria. Clinical follow-ups were performed at twelve months. The endpoints were the occurrence of all-cause death, definite stent thrombosis, and new revascularization. Results. Between November 2017 and May 2019, 790 patients were included (1067 lesions). The mean age was 60.42 ± 14.94 years, and 74.7% presented with acute coronary syndrome. Diabetes mellitus was present in 43.9% of patients, and previous myocardial infarction and previous percutaneous coronary intervention were present in 17.9% and 11.3%, respectively. Angiographic success was achieved in 99.1%. The incidence of all-cause death was 11.5% (6.2% in-hospital and 5.3% in the follow-up) and definitive stent thrombosis was 0.2%. New revascularization was performed in only 5.8% (target lesion revascularization: 2.2%; progression of disease in another lesion: 3.6%). Based on the multivariate regression analysis, only chronic renal failure was an independent predictor of adverse events (OR: 3.3; 95% CI: 1.22–8.92). Conclusion. The result of this single-center registry demonstrates the safety and excellent performance of the InspironTM stent in daily clinical practice with a low rate of adverse cardiac events.

Publisher

Hindawi Limited

Subject

Cardiology and Cardiovascular Medicine,Radiology, Nuclear Medicine and imaging

Cited by 2 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3