Affiliation:
1. Analytical Science Team, Quality of life Division , National Physical Laboratory, Teddington, Middlesex TW11 0LW, UK
Abstract
No measurement result of any kind is complete without an associated statement of uncertainty. In general, the aim is to minimise this measurement uncertainty to the extent that the measurement made is fit for purpose, or the uncertainty meets an acceptable level such as a ‘‘data quality objective’’. This paper demonstrates how the sampling time used to measure a pollutant in ambient air over a given time period, particularly with automatic instruments, can affect the uncertainty of the measurement result. It is shown that the uncertainty of the overall result depends critically on how the precision characteristics of the measuring instrumentation vary with analyte abundance. An example of the determination of these characteristics for mercury vapour measurement and the conclusions that may be drawn about optimum sampling times to minimise uncertainty are presented. For situations where uncertainty is minimised by the use of long sampling periods, the implications of this strategy on the ‘‘information richness’’ of the data and on the detection of short-term episodes, as well as peak concentrations, are discussed.
Subject
Computer Science Applications,Instrumentation,Chemical Engineering (miscellaneous),Analytical Chemistry