Abstract
pH is frequently measured in laboratories, but to have confidence in the results it is necessary to know that it was measured properly. For an electrode to give accurate results it must be treated well and calibrated correctly. In this paper, an automated system for pH measurement is described; the system uses the operational pH scale and calibrates using two or three buffer solutions, taking proper account of the effects of temperature on the system. The system can be programmed with standard methods and procedures to ensure that the electrode gives the best possible performance. Calibrations and measurements within the system are reproducible, and the automated system is more robust than the manual pH meter, and requires less operator time.