Affiliation:
1. School of Earth, Ocean and Environmental Sciences, University of Plymouth, Plymouth PL4 8AA, UK
Abstract
An automated flow injection system for the determination of dissolved silver at ultratrace concentrations in seawater, and controlled under LabVIEW, is described. The flow injection system allows online processing of seawater samples before their analysis using a magnetic sector inductively coupled plasma mass spectrometry (MS-ICP-MS) instrument. Samples were analysed with a minimum amount of manipulation, thereby reducing the risk of contamination. In addition, the flow injection approach with incorporation of an anion exchange minicolumn allowed ready removal of analytical interferences caused by the saline matrix. The software allowed full control of all flow injection components (valves and pumps) and removed manual time control and, therefore, operator errors. The optimized system was capable of five sample injections per h, including preconcentration and wash steps. The limit of detection was 0.5 pM for a 240-s sample load time, which allowed the determination of dissolved silver in open ocean waters, where picomolar concentration levels are typically encountered.
Subject
Computer Science Applications,Instrumentation,Chemical Engineering (miscellaneous),Analytical Chemistry
Cited by
10 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献