Feasibility of Attachable Ring Stimulator for Intraoperative Neuromonitoring during Thyroid Surgery

Author:

Kim Jongjin1ORCID,Moon Hyeon Jong2,Chai Young Jun1ORCID,Lee Jung-Man3,Hwang Ki-Tae1,Wu Che-Wei4,Dionigi Gianlorenzo5ORCID,Kim Hoon Yub6,Park Kyung Sik7,Kim Sang Wan8ORCID,Yi Ka Hee8

Affiliation:

1. Department of Surgery, Seoul Metropolitan Government-Seoul National University Boramae Medical Center, Seoul 07061, Republic of Korea

2. Department of Thoracic and Cardiovascular Surgery, Seoul Metropolitan Government-Seoul National University Boramae Medical Center, Seoul 07061, Republic of Korea

3. Department of Anesthesiology and Pain Medicine, Seoul Metropolitan Government-Seoul National University Boramae Medical Center, Seoul 07061, Republic of Korea

4. Department of Otorhinolaryngology–Head and Neck Surgery, Kaohsiung Medical University Hospital, Kaohsiung Medical University, Kaohsiung, Taiwan

5. Division for Endocrine and Minimally Invasive Surgery, Department of Human Pathology in Adulthood and Childhood “G. Barresi”, University Hospital “G. Martino”, University of Messina, Messina, Italy

6. Department of Surgery, KUMC Thyroid Center, Korea University Anam Hospital, Korea University College of Medicine, Seoul 02841, Republic of Korea

7. Department of Surgery, Konkuk University School of Medicine, Seoul 05030, Republic of Korea

8. Department of Internal Medicine, Seoul National University College of Medicine and Seoul Metropolitan Government-Seoul National University Boramae Medical Center, Seoul 07061, Republic of Korea

Abstract

Objective. Stimulator-attached dissecting instruments are useful for intraoperative nerve monitoring during thyroidectomy. The aim of this study was to evaluate the feasibility of an attachable ring stimulator (ARS) by comparing the electromyography (EMG) amplitudes evoked by an ARS and a conventional stimulator. Methods. Medical records of fourteen patients who underwent thyroidectomy using intraoperative neuromonitoring between June and August 2019 were retrospectively reviewed. The amplitudes of V1, R1, R2, and V2 signals were checked using both the ARS and a conventional stimulator, at the same point. Results. Both stimulators were tested on 20 recurrent laryngeal nerves (RLNs) and 20 vagus nerves (VNs). In all the nerves, the amplitudes of V1, R1, R2, and V2 were greater than 500 μV. The mean amplitudes of V1, R1, R2, and V2 checked with the ARS were 1175, 1432, 1598, and 1279 μV, respectively. The mean amplitudes of V1, R1, R2, and V2 checked with the conventional stimulator were 1140, 1425, 1557, and 1217 μV, respectively. Difference between amplitudes evoked by the two stimulators for V1, R1, R2, and V2 was 77, 110, 102, and 99 μV, respectively. There was no statistical difference in the amplitudes between the two groups for V1, R1, R2, and V2. Conclusion. The ARS transferred electric stimulation as effectively as the conventional stimulator. It is an effective tool for repeated stimulation and facilitates continuous feedback regarding the functional integrity of nerves during thyroid surgery.

Funder

National Research Foundation of Korea

Publisher

Hindawi Limited

Subject

Endocrine and Autonomic Systems,Endocrinology,Endocrinology, Diabetes and Metabolism