Affiliation:
1. Department of Mechanical Engineering, McGill University, MacDonald Engineering Building, 817 Rue Sherbrooke Ouest #270, Montréal, Québec, H3A 0C3, Canada
Abstract
Most surgical simulators leverage virtual or bench models to simulate reality. This study proposes and validates a method for workspace configuration of a surgical simulator which utilizes a haptic device for interaction with a virtual model and a bench model to provide additional tactile feedback based on planned surgical manoeuvers. Numerical analyses were completed to determine the workspace and position of a haptic device, relative to the bench model, used in the surgical simulator, and the determined configuration was validated using device limitations and user data from surgical and nonsurgical users. For the validation, surgeons performed an identical surgery on a cadaver prior to using the simulator, and their trajectories were then compared to the determined workspace for the haptic device. The configuration of the simulator was determined appropriate through workspace analysis and the collected user trajectories. Statistical analyses suggest differences in trajectories between the participating surgeons which were not affected by the imposed haptic workspace. This study, therefore, demonstrates a method to optimally position a haptic device with respect to a bench model while meeting the manoeuverability needs of a surgical procedure. The validation method identified workspace position and user trajectory towards ideal configuration of a mixed reality simulator.
Subject
General Immunology and Microbiology,General Biochemistry, Genetics and Molecular Biology,General Medicine
Cited by
3 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献