Contrast Microsphere Destruction by a Continuous Flow Ventricular Assist Device: An In Vitro Evaluation Using a Mock Circulation Loop

Author:

Platts David G.123ORCID,Bartnikowski Nicole4,Gregory Shaun D.235ORCID,Scalia Gregory M.126,Fraser John F.237

Affiliation:

1. Department of Echocardiography, The Prince Charles Hospital, Brisbane, QLD, Australia

2. School of Medicine, University of Queensland, Brisbane, QLD, Australia

3. Innovative Cardiovascular Engineering and Technology Laboratory, Critical Care Research Group, The Prince Charles Hospital, Brisbane, QLD, Australia

4. School of Chemistry, Physics and Mechanical Engineering, Queensland University of Technology, Brisbane, QLD, Australia

5. School of Engineering, Griffith University, Brisbane, QLD, Australia

6. Heart Care Partners, Wesley Hospital, Brisbane, QLD, Australia

7. Adult Intensive Care Service, The Prince Charles Hospital, Brisbane, QLD, Australia

Abstract

Objectives. Transthoracic echocardiography (TTE) is fundamental in managing patients supported with ventricular assist devices (VAD). However imaging can be difficult in these patients. Contrast improves image quality but they are hydrodynamically fragile agents. The aim was to assess contrast concentration following passage through a VAD utilising a mock circulation loop (MCL). Methods. Heartware continuous flow (CF) VAD was incorporated into a MCL. Definity® contrast was infused into the MCL with imaging before and after CF-VAD. 5 mm2 regions of interest were used to obtain signal intensity (decibels), as a surrogate of contrast concentration. Results. Four pump speeds revealed significant reduction in contrast signal intensity after CF-VAD compared to before CF-VAD (all p<0.0001). Combined pre- and postpump data at all speeds showed a 22.2% absolute reduction in contrast signal intensity across the CF-VAD (14.8 ± 0.8 dB prepump versus 11.6 ± 1.4 dB postpump; p<0.0001). Mean signal intensity reduction at each speed showed an inverse relationship between speed and relative reduction in signal intensity. Conclusion. Contrast microsphere transit through a CF-VAD within a MCL resulted in significant reduction in signal intensity, consistent with destruction within the pump. This was evident at all CF-VAD pump speeds but relative signal drop was inversely proportional to pump speed.

Publisher

Hindawi Limited

Subject

General Immunology and Microbiology,General Biochemistry, Genetics and Molecular Biology,General Medicine

Cited by 3 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

1. The role of innovative modeling and imaging techniques in improving outcomes in patients with LVAD;Frontiers in Cardiovascular Medicine;2023-08-24

2. Intraventricular Flow Patterns in Patients Treated with Left Ventricular Assist Devices;ASAIO Journal;2020-04-17

3. Contrast Echocardiography in Critical Care;State of the Art Techniques in Critical Care Echocardiography;2019-11-23