Accuracy of Dose Delivery in Multiple Breath-Hold Segmented Volumetric Modulated Arc Therapy: A Static Phantom Study

Author:

Noto Kimiya1,Ueda Shinichi1,Kojima Hironori1,Isomura Naoki1,Takemura Akihiro2,Takamatsu Shigeyuki3,Kumano Tomoyasu3,Takanaka Tsuyoshi4

Affiliation:

1. Department of Radiology, Kanazawa University Hospital, 13-1 Takara-machi, Kanazawa, Ishikawa 920-8641, Japan

2. Faculty of Health Sciences, Institute of Medical Pharmaceutical and Health Sciences, Kanazawa University, 5-11-80 Kodatsuno, Kanazawa, Ishikawa 920-0942, Japan

3. Department of Radiotherapy, Graduate School of Medical Science, Kanazawa University, 13-1 Takara-machi, Kanazawa, Ishikawa 920-8641, Japan

4. Department of Radiation Oncology, Kouseiren Takaoka Hospital, 5-10 Eirakucho, Takaoka, Toyama 933-8555, Japan

Abstract

Purpose. Accuracy of dose delivery in multiple breath-hold segmented volumetric modulated arc therapy (VMAT) was evaluated in comparison to noninterrupted VMAT using a static phantom. Material and Methods. Five VMAT plans were evaluated. A Synergy linear accelerator (Elekta AB, Stockholm, Sweden) was employed. A VMAT delivery sequence was divided into multiple segments according to each of the predefined breath-hold periods (10, 15, 20, 30, and 40 seconds). The segmented VMAT delivery was compared to noninterrupted VMAT delivery in terms of the isocenter dose and pass rates of a dose difference of 1% with a dose threshold of 10% of the maximum dose on a central coronal plane using a two-dimensional dosimeter, MatriXX Evolution (IBA Dosimetry, Schwarzenbruck, Germany). Results. Means of the isocenter dose differences were 0.5%, 0.2%, 0.2%, 0.0%, and 0.0% for the beam-on-times between interrupts of 10, 15, 20, 30, and 40 seconds, respectively. Means of the pass rates were 85%, 99.9%, 100%, 100%, and 100% in the same order as the above. Conclusion. Our static phantom study indicated that the multiple breath-hold segmented VMAT maintains stable and accurate dose delivery when the beam-on-time between interrupts is 15 seconds or greater.

Publisher

Hindawi Limited

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3