Affiliation:
1. Instituto de Química, Universidade Federal de Uberlândia, 38400-902 Uberlândia, MG, Brazil
Abstract
In this study, a method for the determination of low concentrations of lead in beer samples using solid-phase extraction with a flow injection analysis system and detection by flame atomic absorption spectrometry (FAAS) was developed.Moringa oleiferaseeds were used as a biosorbent material. Chemical and flow variables of the online preconcentration system, such as sample pH, preconcentration flow rate, eluent flow rate, eluent concentration, particle size, and sorbent mass, were studied. The optimum extraction conditions were obtained using a sample pH of 6.0, sample flow rate of 6.0 mL min−1, 63.0 mg of sorbent mass, and 2.0 mol L−1HNO3at a flow rate of 2.0 mL min−1as the eluent. With the optimized conditions, the preconcentration factor, precision, detection limit, consumption index, and sample throughput were estimated as 93, 0.3% (10.0 μg L−1,n=7), 7.5 μg L−1, 0.11 mL, and 23 samples per hour, respectively. The method developed was successfully applied to beer samples and recovery tests, with recovery ranging from 80% to 100%.
Funder
Conselho Nacional de Desenvolvimento Científico e Tecnológico
Subject
Computer Science Applications,Instrumentation,Chemical Engineering (miscellaneous),Analytical Chemistry