Affiliation:
1. Tianjin Key Laboratory of Marine Resources and Chemistry, Modern Analytical Technology Research Center, College of Chemical Engineering and Materials Science, Tianjin University of Science and Technology, Tianjin 300457, China
Abstract
A sensitive and accurate simultaneous continuous analysis for six arsenic species including arsenobetaine (AsB), arsenocholine (AsC), monomethylarsonic acid (MMA), dimethylarsinic acid (DMA), arsenite (AsIII), and arsenate (AsV) has been developed by high performance liquid chromatography with inductively coupled plasma mass spectrometry (HPLC-ICP-MS). An anion-exchange column of Hamilton PRP-X100 (Switzerland) was applied for separation of the six arsenic species with gradient elution of 1.25 mmol/L Na2HPO4 and 11.0 mmol/L KH2PO4 as the mobile phase A and 2.5 mmol/L Na2HPO4 and 22.0 mmol/L KH2PO4 as the mobile phase B. The linearity ranges for AsB, AsC, MMA, DMA, AsIII, and AsV were between 0.5 and 50.0 μg/L, and the detection limits of the six arsenic species were all within 0.01–0.35 ng/L. The relative standard deviations (RSDs) were within 2.26–3.68% and the recovery rates of samples ranged from 95 to 103%. The proposed method was applied for the arsenic speciation analysis of sediment pore-water samples, which were taken from the supernatant after centrifugation and filtration.
Funder
National Natural Science Foundation of China
Cited by
12 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献