Author:
Nagarajan R.,Gupta A.,Mehrotra R.,Bajaj M. M.
Abstract
Mid-infrared (MIR) spectroscopy in attenuated total reflectance (ATR) mode was used for quantifying ethanol, sucrose, and tartaric acid in alcoholic beverages. One hundred synthetic samples were prepared with different ethanol, sucrose, and tartaric acid concentrations. Experiments were carried out on Bio-Rad 175 C FTS using an ATR accessory. Spectra were recorded in the wavelength region 600–4000cm−1. Calibration was performed using partial least squares (PLS) algorithm. Commercially available alcoholic beverages (gin, rum, vodka, etc.) were experimented and concentration of ethanol in these samples was predicted using the developed calibration model. Chemical analysis of these commercial samples was carried out in order to compare the results. The agreement between ATR results with those of chemical analysis revealed good reliability and repeatability of the technique used.
Funder
Council of Scientific and Industrial Research
Subject
Computer Science Applications,Instrumentation,Chemical Engineering (miscellaneous),Analytical Chemistry
Cited by
16 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献