Biomechanical Comparison between Isobar and Dynamic-Transitional Optima (DTO) Hybrid Lumbar Fixators: A Lumbosacral Finite Element and Intersegmental Motion Analysis

Author:

Chen Shih-Hao1,Hsiao Chih-Kun2,Wang Chih-Wei1,Chen Hsiang-Ho3,Zhong Zheng-Cheng4ORCID

Affiliation:

1. Department of Orthopaedics, Tzu Chi General Hospital at Dalin and Tzu Chi University, Taiwan

2. Department of Orthopedics, E-Da Hospital, Kaohsiung, Taiwan

3. Center for Biomedical Engineering, Department of Biomedical Engineering, College of Engineering, Chang Gung University, Taoyuan 33302, Taiwan

4. Department of Mechanical Engineering, National Yang Ming Chiao Tung University, Hsinchu, Taiwan

Abstract

Biomechanical performance of longitudinal component in dynamic hybrid devices was evaluated to display the load-transfer effects of Dynesys cord spacer or Isobar damper-joint dynamic stabilizer on junctional problem based on various disc degenerations. The dynamic component was adapted at the mildly degenerative L3–L4 segment, and the static component was fixed at the moderately degenerative L4–L5 segment under a displacement-controlled mode for the finite element study. Furthermore, an intersegmental motion behavior was analyzed experimentally on the synthetic model under a load-controlled mode. Isobar or DTO hybrid fixator could reduce stress/motion at transition segment, but compensation was affected at the cephalic adjacent segment more than the caudal one. Within the trade-off region (as a motion-preserving balance between the transition and adjacent segments), the stiffness-related problem was reduced mostly in flexion by a flexible Dynesys cord. In contrast, Isobar damper afforded the effect of maximal allowable displacement (more than peak axial stiffness) to reduce stress within the pedicle and at facet joint. Pedicle-screw travel at transition level was related to the extent of disc degeneration in Isobar damper-joint (more than Dynesys cord spacer) attributing to the design effect of axial displacement and angular rotation under motion. In biomechanical characteristics relevant to clinical use, longitudinal cord/damper of dynamic hybrid lumbar fixators should be designed with less interface stress occurring at the screw-vertebral junction and facet joint to decrease pedicle screw loosening/breakage under various disc degenerations.

Publisher

Hindawi Limited

Subject

General Immunology and Microbiology,General Biochemistry, Genetics and Molecular Biology,General Medicine

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3