Affiliation:
1. Atomic Institute, Vienna University of Technology, Stadionallee 2, 1020 Vienna, Austria
Abstract
A fully automated and fast pneumatic transport system for short-time activation analysis was recently developed. It is suitable for small nuclear research reactors or laboratories that are using neutron generators and other neutron sources. It is equipped with a programmable logic controller, software package, and 12 devices to facilitate optimal analytical procedures. 550 ms were only necessary to transfer the irradiated capsule (diameter: 15 mm, length: 50 mm, weight: 4 gram) to the counting chamber at a distance of 20 meters using pressurized air (4 bars) as a transport gas.
Subject
Computer Science Applications,Instrumentation,Chemical Engineering (miscellaneous),Analytical Chemistry
Cited by
12 articles.