Corpuls CPR Generates Higher Mean Arterial Pressure Than LUCAS II in a Pig Model of Cardiac Arrest

Author:

Eichhorn S.1ORCID,Mendoza A.2,Prinzing A.1,Stroh A.1,Xinghai L.1,Polski M.1,Heller M.3,Lahm H.1,Wolf E.4,Lange R.15,Krane M.15

Affiliation:

1. Institute for Translational Cardiac Surgery (INSURE), Department of Cardiovascular Surgery, German Heart Center Munich, Technische Universität München, Munich, Germany

2. Fakultät für Informatik, Robotics and Embedded Systems, Technische Universität München, Munich, Germany

3. GS Elektromedizinische Geräte GmbH, Kaufering, Germany

4. Gene Center, LMU Munich, Munich, Germany

5. German Center for Cardiovascular Research (DZHK), Partner Site Munich Heart Alliance, Munich, Germany

Abstract

According to the European Resuscitation Council guidelines, the use of mechanical chest compression devices is a reasonable alternative in situations where manual chest compression is impractical or compromises provider safety. The aim of this study is to compare the performance of a recently developed chest compression device (Corpuls CPR) with an established system (LUCAS II) in a pig model. Methods. Pigs (n = 5/group) in provoked ventricular fibrillation were left untreated for 5 minutes, after which 15 min of cardiopulmonary resuscitation was performed with chest compressions. After 15 min, defibrillation was performed every 2 min if necessary, and up to 3 doses of adrenaline were given. If there was no return of spontaneous circulation after 25 min, the experiment was terminated. Coronary perfusion pressure, carotid blood flow, end-expiratory CO2, regional oxygen saturation by near infrared spectroscopy, blood gas, and local organ perfusion with fluorescent labelled microspheres were measured at baseline and during resuscitation. Results. Animals treated with Corpuls CPR had significantly higher mean arterial pressures during resuscitation, along with a detectable trend of greater carotid blood flow and organ perfusion. Conclusion. Chest compressions with the Corpuls CPR device generated significantly higher mean arterial pressures than compressions performed with the LUCAS II device.

Funder

Bayerische Forschungsstiftung

Publisher

Hindawi Limited

Subject

General Immunology and Microbiology,General Biochemistry, Genetics and Molecular Biology,General Medicine

Cited by 9 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3