Affiliation:
1. Faculty of Medicine, University of Banja Luka, 78000 Banja Luka, Bosnia and Herzegovina
2. Faculty of Chemistry, University of Belgrade, 11000 Belgrade, Serbia
Abstract
This work presents the development of a flow injection system for differential pulse amperometry (DPA) for diazepam determination in the presence of oxygen. The thin flow cell consisted of the bare glassy carbon electrode, reference silver/silver chloride, and stainless steel as the auxiliary electrode. Electrochemical reduction of diazepam (DZP) was characterised by cyclic voltammetry. Azomethine reduction peak was used for DZP quantification. The detector response was linear in the range 20–250 µmol/dm3of diazepam, with a calculated detection limit of 3.83 µg/cm3. Intraday and interday precision were 1.53 and 10.8%, respectively. The method was applied on three beverage samples, energetic drink, and two different beer samples, and obtained recoveries were from 93.65 up to 104.96%. The throughoutput of the method was up to 90 analyses per hour.
Funder
Ministarstvo Prosvete, Nauke i Tehnološkog Razvoja
Subject
Computer Science Applications,Instrumentation,General Chemical Engineering,Analytical Chemistry
Cited by
6 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献