Quantitative Analysis of Favipiravir by HPLC: Development and Validation

Author:

Qurt Moammal1ORCID,Eshtayyeh Rania1,Naseef Hani1ORCID,Rabba Abdullah1ORCID,Abukhalil Abdallah Damin1ORCID,Malkieh Numan2ORCID,Kharouf Maher2

Affiliation:

1. Pharmacy Department, Faculty of Pharmacy, Nursing and Health Professions, Birzeit University, P.O. Box 14, West Bank, Birzeit, Palestine, State of Palestine

2. Jerusalem Pharmaceuticals, Al Bireh, P.O. Box 3570, West Bank, Ramallah, State of Palestine

Abstract

Favipiravir is a broad-spectrum oral antiviral agent and has been approved for the treatment of COVID-19 infection cases. It inhibits a protein known as RNA polymerase, which transcribes and replicates the viral RNA genome, causing the spread of the infection. The current study aimed to develop and validate a new analytical method utilizing HPLC in accordance with international requirements (ICH and FDA). The chromatographic conditions used to achieve good resolution and reproducibility were a mixture of acetonitrile and 0.1% phosphoric acid buffer in the ratio of 60 : 40 v/v as the mobile phase. The flow rate was 1.0 mL/min, the wavelength (λ) was determined at 250 nm, and a retention time was approximately 3 minutes for favipiravir. The HPLC analysis was performed on the Dionex 300 system equipped with a Phenomenex C8 (250 cm 4.6 mm) 5 μm column. The total runtime was 6.0 min. The findings indicated that the method had been validated satisfactorily. Across the concentration range of 0.10–0.75 mg/ml, the calibration curve revealed a linear relationship. The accuracy of the current method was to be 99.2%. The limit of detection (LOD) and limit of quantification (LOQ) were 0.004 and 0.013 ppm, respectively. The standard and sample solution repeatability tests revealed that the procedure was precise and within acceptable ranges. The RSD% for the determination of precision was <2%. The results for robustness and solution stability were within acceptable limits. Finally, the new method provided an excellent result for all analytical method validation parameters and met the acceptance criteria. In addition, the new approach has a short run time and a retention time of around 4 minutes.

Publisher

Hindawi Limited

Subject

Analytical Chemistry

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3