Development and Application of a Standardized Testset for an Artificial Intelligence Medical Device Intended for the Computer-Aided Diagnosis of Diabetic Retinopathy

Author:

Wang Hao1ORCID,Meng Xiangfeng1ORCID,Tang Qiaohong1ORCID,Hao Ye1ORCID,Luo Yan2ORCID,Li Jiage1ORCID

Affiliation:

1. Institute for Medical Device Control, National Institutes for Food and Drug Control, 31 Huatuo Rd, Beijing 102629, China

2. State Key Laboratory of Ophthalmology, Image Reading Center, Zhongshan Ophthalmic Center, Sun Yat-Sen University, No. 54 Xianlie South Road, Yuexiu District, Guangzhou 510060, Guangdong, China

Abstract

Objective. To explore a centralized approach to build test sets and assess the performance of an artificial intelligence medical device (AIMD) which is intended for computer-aided diagnosis of diabetic retinopathy (DR). Method. A framework was proposed to conduct data collection, data curation, and annotation. Deidentified colour fundus photographs were collected from 11 partner hospitals with raw labels. Photographs with sensitive information or authenticity issues were excluded during vetting. A team of annotators was recruited through qualification examinations and trained. The annotation process included three steps: initial annotation, review, and arbitration. The annotated data then composed a standardized test set, which was further imported to algorithms under test (AUT) from different developers. The algorithm outputs were compared with the final annotation results (reference standard). Result. The test set consists of 6327 digital colour fundus photographs. The final labels include 5 stages of DR and non-DR, as well as other ocular diseases and photographs with unacceptable quality. The Fleiss Kappa was 0.75 among the annotators. The Cohen’s kappa between raw labels and final labels is 0.5. Using this test set, five AUTs were tested and compared quantitatively. The metrics include accuracy, sensitivity, and specificity. The AUTs showed inhomogeneous capabilities to classify different types of fundus photographs. Conclusions. This article demonstrated a workflow to build standardized test sets and conduct algorithm testing of the AIMD for computer-aided diagnosis of diabetic retinopathy. It may provide a reference to develop technical standards that promote product verification and quality control, improving the comparability of products.

Funder

National Basic Research Program of China

Publisher

Hindawi Limited

Subject

Health Informatics,Biomedical Engineering,Surgery,Biotechnology

Cited by 3 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3