The Effect of Mini Dental Implant Number on Mandibular Overdenture Retention and Attachment Wear

Author:

Alshenaiber Rafif1ORCID,Barclay Craig2ORCID,Silikas Nick3ORCID

Affiliation:

1. Prosthetic Dental Sciences Department at College of Dentistry, Prince Sattam Bin Abdulaziz University, Al-Kharj, Saudi Arabia

2. Oral Rehabilitation, University Dental Hospital of Manchester, Manchester, UK

3. Division of Dentistry, Faculty of Biology, Medicine and Health, University of Manchester, Coupland 3 Building, Manchester M13 9PL, UK

Abstract

Purpose. Evaluate the effect of different mini-implant numbers on overdenture retention and evaluate attachment wear following one year of simulated placement/removal. Material and Methods. Nine models simulating atrophic mandibles held 27 mini dental implants in three groups of 2, 3, and 4 mini-implants. A total of 1080 simulated placement/removal cycles were carried out, and a digital force gauge was used to measure the overdenture dislodgment force. The means of the retention forces were analyzed using SPSS with one-way ANOVA and post hoc ( p < 0.05 ). The inner diameter of attachment inserts was evaluated using a light microscope before and after testing. A paired t -test was used to compare the mean of inner ring diameters ( p < 0.05 ). Results. The retention was significantly reduced regardless of the mini dental implant number, but the number affected overdenture retention. The placement of 4 mini dental implants provided higher retention and less reduction in retentiveness. However, no significant difference was found when 3 mini dental implants were compared to 2 mini dental implants ( p = 0.21 ). Microscopic examination showed abrasion wear in all inserts following testing. However, the inserts of the 4 mini dental implants showed less wear than those used for 2 or 3 mini dental implants with p 0.001 and p 0.001 , respectively. Conclusion. Mini dental implant overdenture retention force and attachment wear could improve by increasing the mini dental implants to 4. However, there was no difference in retention force or attachment wear when 2 or 3 mini dental implant overdentures were compared.

Funder

Saudi Cultural Bureau

Publisher

Hindawi Limited

Subject

General Immunology and Microbiology,General Biochemistry, Genetics and Molecular Biology,General Medicine

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3